Drager Babylog VN500

Preventive Maintenance Procedure

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Babylog VN500

Manual Reference

Infinity Acute Care System – Babylog VN500 Ventilation Unit Instructions for Use, Software 2.n, Maintenance chapter, pages 207–210.

Manufacturer Recommendation / AEM Status

The manufacturer-listed inspection and safety-inspection interval is every year and must be performed by trained service personnel.

Manufacturer-Listed Maintenance Intervals

Every 4 Weeks

Clean the Babylog VN500 ambient air filter and exchange the filter when necessary. The manual directs cleaning or replacement after four weeks at the latest.

Every 12 Months

Complete the following:

Every Year

Check the capacity of the internal Babylog VN500 battery and the PS500 internal batteries when installed. Replace batteries sooner when the capacity check indicates that replacement is necessary.

Every 2 Years at the Latest

Complete the following:

Every 6 Years

Replace the filter in the O₂ gas inlet.

The Air and O₂ gas-inlet filters, battery replacement, GS500 filter service, device inspection and service, and safety inspection are assigned to trained service personnel.

No replacement part numbers for the ambient air filter, expiratory-valve diaphragm, batteries, GS500 filters, or gas-inlet filters are provided in this manual.

Manufacturer-Listed Safety Inspection

The annual safety inspection includes:

The manufacturer states that the safety inspection does not replace the listed inspection, maintenance, or preventive replacement of wearing parts.

Post-Repair Safety Inspection

The provided manual states that the device must be inspected and serviced regularly by trained service personnel, but it does not provide a separate universal post-repair test matrix.

After repair, perform all safety, functional, calibration, device-check, breathing-circuit-check, and electrical-safety activities applicable to the component or system disturbed by the repair. Follow the applicable service documentation before returning the ventilator to clinical use.

AEM Note

If a facility performs this PM at a different interval or changes the listed activities, the change should be managed through the facility’s approved Alternative Equipment Maintenance program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Use calibrated equipment suitable for the installed Babylog VN500 configuration.

The provided manual does not specify part numbers for the scheduled filter, diaphragm, battery, GS500 filter, or gas-inlet-filter replacements.

Step-by-Step PM Procedure

1. Remove the Ventilator From Clinical Use

Action

Confirm that:

Expected outcome: Maintenance can be completed without affecting patient care or exposing service personnel to contamination.

2. Identify the Installed Configuration

Action

Record:

Confirm which interval-based replacements are due.

Expected outcome: All applicable components and maintenance intervals are included.

3. Inspect Documentation and Equipment Completeness

Action

Verify:

Expected outcome: The ventilator configuration is complete and supported by the available documentation.

4. Perform the External Visual Inspection

Action

Inspect:

Confirm:

Expected outcome: No physical condition exists that could impair safe operation.

5. Service the Babylog VN500 Ambient Air Filter

Action

The manufacturer-listed interval is cleaning or exchange after four weeks at the latest and replacement after one year.

  1. Unscrew the ambient-air-filter cover screw.
  2. Open the cover.
  3. Remove the filter.
  4. Install a new filter when the annual replacement is due or the existing filter is damaged.
  5. For interim cleaning, wash the reusable filter in warm soapy water.
  6. Dry the filter thoroughly.
  7. Insert the filter without folding or creasing it.
  8. Close the cover and tighten the screw.

Expected outcome: The filter is clean, dry, correctly installed, and within its manufacturer-listed replacement interval.

6. Replace the Expiratory-Valve Diaphragm

Action

  1. Replace the diaphragm annually.
  2. Remove the expiratory valve according to the Instructions for Use.
  3. Remove the existing diaphragm.
  4. Inspect the valve housing for contamination or damage.
  5. Fit the new diaphragm evenly around the edge of the expiratory-valve housing.
  6. Confirm that the diaphragm is seated correctly without folds or distortion.
  7. Reinstall and lock the expiratory valve.
  8. Confirm secure engagement by gently pulling on the valve.

Expected outcome: The new diaphragm is correctly seated and the expiratory valve is securely installed.

7. Complete Other Interval-Based Service

Action

Complete or verify completion of the following when applicable:

Battery, GS500, and gas-inlet-filter service must be completed by personnel qualified for the applicable service procedure.

Expected outcome: All manufacturer-listed scheduled components are current and documented.

8. Power and Battery Inspection

Action

Connect the system to AC power and verify:

Perform the manufacturer service-level battery capacity and functional tests for:

The manual identifies at least 30 minutes of operation from a fully charged, new Babylog VN500 battery and at least 100 minutes from new, fully charged PS500 batteries under typical ventilation conditions. These stated runtimes describe new-battery performance and should not be treated as a substitute for the manufacturer service battery test.

Expected outcome: AC operation, charging, battery transfer, and battery capacity meet the applicable manufacturer service criteria.

9. Perform the Device Check

Action

Place the ventilator in Standby and run the device check.

The available check sequence may include:

Repeat any failed or incomplete test after correcting the identified cause.

Expected outcome: All applicable test steps display a correct result.

Do not return the ventilator to service when the pressure-sensor calibration valve, expiratory valve, or safety-valve test fails. The manual identifies these as safety-relevant failures and states not to start ventilation.

10. Perform the Breathing-Circuit Check

Action

Using the selected breathing circuit, humidifier configuration, and patient category:

  1. Start the breathing-circuit check.
  2. Follow the displayed instructions to seal and open the Y-piece.
  3. Record leakage.
  4. Record compliance.
  5. Record inspiratory resistance.
  6. Record expiratory resistance.

Manufacturer-Listed Leakage Criterion

At 60 mbar or 60 cmH₂O, leakage must be no more than 300 mL/min.

Expected outcome: Leakage is within the listed limit and circuit compliance and resistance are successfully measured.

11. Verify Safety Devices

These tests require appropriate calibrated flow- and pressure-measurement equipment and the applicable service procedure.

Emergency Expiratory Valve

Action: Apply a flow of 4.5 to 5.5 L/min and measure the resulting pressure rise.

Expected outcome: Pressure rise is 1.9 to 4.4 mbar, equivalent to 1.9 to 4.4 cmH₂O.

Expiratory-Valve Non-Return Valve

Action: Perform the manufacturer service test for correct one-way-valve operation.

Expected outcome: The valve prevents reverse flow as specified.

Emergency Breathing Valve

Action: Apply a suction flow of 60 to 65 L/min and measure the pressure drop.

Expected outcome: Maximum pressure drop is 6 mbar, equivalent to 6 cmH₂O.

Gas-Inlet Non-Return Valves

Action: Verify correct operation of the non-return valves in both the O₂ gas inlet and Air gas inlet.

Expected outcome: Both valves operate correctly without prohibited reverse flow.

The safety-device values above are manufacturer-listed annual safety-inspection criteria.

12. Verify Airway-Pressure and Flow Measurement

Action

Using appropriate calibrated test equipment:

Expected outcome: Both measurement systems meet the manufacturer service specifications.

Manual limitation: The provided Instructions for Use list these tests but do not provide the complete test setup or numerical accuracy limits. Do not invent tolerances.

13. Verify the Alarm System

Action

Verify:

Expected outcome: Local visual and audible alarms operate reliably, and the installed remote alarm output functions correctly.

Electrical Safety Testing

The annual manufacturer-listed safety inspection requires electrical safety testing according to IEC 62353 or DIN VDE 0751, Part 1.

Test the complete installed device combination as applicable, including the Babylog VN500, Infinity C500, PS500, GS500, and powered accessories.

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:

Result =

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Document the standard, test method, device configuration, and limits used.

The Babylog VN500 manual does not provide a specific ground-resistance or leakage-current limit. Do not substitute an invented manufacturer limit.

15. Complete Final Functional Check

Action

After completing maintenance:

Expected outcome: The system is complete, correctly configured, and ready for clinical setup.

Pass / Fail Criteria

Pass

The ventilator passes PM when:

Remove From Service

Remove the ventilator from service when:

Copy/Paste CMMS Work Order Procedure

Scheduled PM

Filter Change Only

Post-Repair Safety Inspection

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide