GE Healthcare Aisys CS2

Preventive Maintenance Procedure

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Aisys CS2

Manual Reference

Aisys CS2 Technical Reference Manual, 2069137-001, Revision F, August 2024.
Aisys CS2 Planned Maintenance Checklist, 2069137-PMC, Revision E, August 2024.

Manufacturer Recommendation / AEM Status

GE Healthcare lists the main planned-maintenance procedure at a 12-month interval, except for activities specifically assigned a different interval. The manufacturer describes this as a sequential maintenance workflow for trained personnel and states that parts due for replacement should be replaced before completing the checks, tests, and calibrations.

Flow and pressure test equipment should have an accuracy of ±2.5% of reading or better unless a specific test states otherwise.

Manufacturer-Listed Replacement Intervals

Every 12 Months

Every 24 Months

Every 48 Months

The manufacturer checklist includes a field to identify the applicable 6-, 12-, 24-, or 48-month service interval. The standard functional PM activities are listed at 12 months, while the 24- and 48-month selections identify additional interval-based replacement work.

Respiratory gas modules require maintenance under the applicable gas-module Technical Reference Manual. The Aisys CS2 procedure does not replace the separate module-specific maintenance procedure.

If the facility performs this maintenance at a different interval or changes the manufacturer-listed activities, the change should be managed through the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Do not perform testing or maintenance while the system is being used on a patient. Follow facility infection-control procedures because used equipment may contain blood or body fluids.

Step-by-Step PM Procedure

1. Verify Equipment Status and Service Interval

Action:

Expected outcome: The correct PM scope and interval-based replacements are identified before testing begins.

2. Replace Interval-Based Parts

Action: Replace applicable manufacturer-listed parts before completing the functional tests.

Follow the detailed replacement procedures in the Technical Reference Manual. Erase or manage logs as specified when replacing the display battery.

Expected outcome: All parts due at the applicable service interval are replaced and documented.

3. Perform the Visual and Mechanical Inspection

Action: Inspect the following:

Clean reusable filters when permitted by facility procedure or replace damaged filters as appropriate.

Expected outcome: Filters are clean; structural parts are secure and undamaged; casters roll and lock properly; labels are legible; hoses, tubing, and cables show no deterioration; and the display arm remains positioned and moves correctly.

4. Verify Startup Annunciators

Action: Turn the system switch on and listen during startup.

Expected outcome: All four manufacturer-listed startup audible indications occur.

5. Test Pipeline and Cylinder Supplies

Action:

Expected outcome: Each pipeline and cylinder is correctly identified. With supplies disconnected, gauges read zero and the low-O₂ alarm occurs. Cylinder pressure decay after one minute is less than 690 kPa / 100 psi for each cylinder.

Record each cylinder separately.

6. Perform the O₂ Supply Alarm Test

Action: Perform the manufacturer-listed O₂ supply alarm test.

Expected outcome:

7. Inspect and Test the AGSS

Action:

Expected outcome: Scavenging components are connected, unobstructed, undamaged, and mechanically functional. Test pressure does not rise by more than 10 cmH₂O.

8. Inspect the Advanced Breathing System

Action: Inspect the following:

Perform the bellows-drop test with the assembly upright and inverted.

Expected outcome: Components are clean, correctly assembled, undamaged, and functional. The bellows does not fall within one minute in either the upright or inverted test position.

9. Inspect the Free-Breathing Valve

Action:

Expected outcome: The free-breathing valve is correctly installed, moves freely, and passes the applicable ventilator checks.

10. Review Logs and Run Ventilator Tests

Action:

Expected outcome:

Do not adjust the machine solely to obtain a passing result without investigating the cause of an initial failure.

11. Run Display Diagnostics

Action:

Expected outcome: LEDs flash five times; the speaker sounds for two seconds; the battery test displays Passed; the speaker sounds as each hard key is pressed; the LCD cycles through the test screens; and touch targets calibrate successfully.

12. Test Network Connectivity

This step applies to System Software 11 and 12.

Action: Run the Network Connectivity Test for all configured ports and devices.

Expected outcome: Each enabled interface shows a green Pass indication. A red X indicates a configuration problem, and an orange question mark indicates a runtime error. Unconfigured interfaces may show Not Applicable.

Document results for Port 1, the Port 1 device, Port 2, Port 2 devices, and the log-transfer interface. Network testing is only applicable to configured interfaces.

13. Test the Electronic Vaporizer

Action:

Expected outcome: The test-cassette procedure displays Pass or Complete. Therapy-cassette leak flow is stable and ≤10 ml/min.

Remove from service any cassette that leaks or is not recognized.

14. Test the Electronic Gas Mixer

Mixer Outlet Check-Valve Leak Test

Action: Pressurize the test setup to 200 mmHg and observe pressure for 30 seconds.

Expected outcome: Pressure decreases by no more than 10 mmHg in 30 seconds.

Mixer Flow Verification

Action: Measure 100% O₂, air, and N₂O at 10 l/min and at 0.5 l/min.

Expected outcome:

15. Test the Alternate O₂ Flowmeter

Complete this step when the alternate O₂ flowmeter is installed.

Action:

Expected outcome:

16. Test the Auxiliary O₂ Flowmeter

Complete this step when the auxiliary O₂ flowmeter is installed.

Action:

Expected outcome:

17. Test the Integrated Suction Regulator

Complete this step when the integrated suction regulator is installed.

Action: Test the gauge zero position, gauge accuracy at 100, 300, and 500 mmHg, maximum flow, regulator setpoint, vacuum bleed, and vacuum leakage with the regulator off.

Expected outcome:

18. Perform the O₂ Flush-Flow Test

Action: Perform the flush-flow test using the manufacturer setup and measure bellows fill time.

Expected outcome: The bellows fills in 1.8–2.3 seconds.

19. Calibrate the Respiratory Gas Module

Complete this step when an applicable airway module is installed.

Action:

Expected outcome: Displayed gas values calibrate to the certified calibration-gas values.

The gas module has its own maintenance requirements. Do not treat this calibration as the complete module PM.

20. Run the System Checkout

Action:

Expected outcome: The Full Test displays Pass or Complete. The SCGO Low P Leak test displays Pass or Complete. For ACGO systems, bulb inflation time is greater than 30 seconds.

21. Perform Alarm Tests

Action: Generate and verify the manufacturer-listed alarms.

Expected outcome: The following alarms occur and are both audible and visible:

Verify that each alarm clears when the test condition is corrected.

22. Test AC Power Failure and Battery Capacity

Action:

Expected outcome: The mains indicator is illuminated on AC power. When AC is disconnected, the message “Plug in power cable. On battery.” appears. The alarm cancels after AC power is restored. The machine operates on battery for more than 45 minutes.

A failed runtime test requires battery investigation or replacement. Do not return the machine to service with unverified backup-power capability.

Electrical Safety Testing

Use an approved medical electrical safety analyzer.

Ground Resistance

Measure between the ground pin of the AC line-cord plug and a pipeline inlet.

Ground resistance measured in ohms, expected ≤ 0.20 Ω:
Result =
Pass / Fail =

The manufacturer’s limit of ≤0.20 Ω is more restrictive than a generic 0.5 Ω facility criterion and should be used for this manufacturer-aligned procedure.

Touch/Chassis Leakage Current

Touch/chassis leakage measured in microamps:
Supply voltage = 100/120 Vac / 220/240 Vac
Facility limit used = Manufacturer limit of <300 µA at 100/120 Vac or <500 µA at 220/240 Vac
Result =
Pass / Fail =

Pass / Fail Criteria

PM Pass Criteria

The Aisys CS2 passes PM when:

Remove From Service Criteria

Remove the machine from service when any of the following remain unresolved:

Final Machine Checks

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide