GE Healthcare Carestation 600 Series

Preventive Maintenance Procedure

On this page

Asset Type

Anesthesia Machine

Manufacturer

GE Healthcare

Model

Carestation 600 Series

Manual Reference

Carestation 620/650/650c (A1/A2) Technical Reference Manual, 2076163-001, Revision U, November 2024.

Carestation 600 Series Planned Maintenance Checklist, 2076164-PMC, Revision 3, November 2024.

The technical reference manual applies to the Carestation 620, 650, and 650c configurations and identifies them collectively as the Carestation 600 Series.

Manufacturer Recommendation / AEM Status

GE provides separate 12-, 24-, 36-, and 48-month planned-maintenance checks. The manufacturer checklist indicates that the complete planned-maintenance inspection and testing sequence is performed at these intervals, with specific parts replaced according to the applicable interval.

Before beginning planned maintenance, GE directs the technician to:

Manufacturer-Listed Parts Replacement

Every 12 Months

Replace the vaporizer port O-rings kit, PN 1102-3016-000.

Every 24 Months

Every 36 Months

Every 48 Months

The system battery is listed as one paired replacement set. The breathing-system check-valve kit quantity is two per system.

AEM Note

If a facility performs this PM at a different interval or changes the listed activities, that change should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Use calibrated equipment appropriate for the installed configuration and the detailed procedures in Technical Reference Manual Section 13.

Do not substitute unverified test equipment or perform internal maintenance without the training, adapters, software access, and procedures required by the technical reference manual.

Step-by-Step PM Procedure

1. Remove the Machine From Clinical Use

Action

Expected outcome: The machine is safely available for maintenance and contamination risks are controlled.

GE states that testing or maintenance must not be performed while the machine is being used on a patient.

2. Determine the Applicable Maintenance Interval

Action

Identify whether this is the 12-month, 24-month, 36-month, or 48-month check. Verify the previous maintenance date and determine which interval-specific parts are due.

Expected outcome: The correct manufacturer-listed replacement parts and maintenance activities are selected.

3. Perform the Initial Full Test and Verify System Time

Action

Expected outcome: The Full Test passes or any preexisting failure is documented before maintenance begins.

Do not continue as though the unit passed if the initial Full Test identifies an unresolved failure.

4. Replace Interval-Specific Parts

Action

Turn off the system before replacing electrical components. Replace all parts due for the selected interval before completing the remaining checks, tests, and calibrations.

Document:

Expected outcome: All manufacturer-listed interval parts are installed correctly.

5. Perform the External Visual and Mechanical Inspection

Action

Inspect:

Expected outcome

Remove the machine from service for damaged gas hoses, deteriorated tubing, compromised electrical wiring, loose structural components, defective brakes, or an unstable frame.

6. Inspect the Breathing System

Action

Complete the detailed technical-reference-manual procedures for:

Perform the bellows-drop test.

Expected outcome

7. Inspect the Ventilator

Action

Inspect the exhalation valve and associated ventilator components using the manual procedure. Replace parts when the inspection identifies deterioration or damage.

Expected outcome: The exhalation valve and ventilator components are correctly assembled, clean, undamaged, and functional.

8. Verify Audible and Visual Annunciators

Action

Expected outcome

9. Perform the Power-Failure and Battery-Capacity Tests

Action

Expected outcome

A battery-runtime result of 50 minutes or less fails the manufacturer checklist.

10. Test Pipeline and Cylinder Gas Supplies

Action

Expected outcome

11. Perform the Flow-Sensing Test

Use the manufacturer setup and test adapter. Record each applicable displayed reading.

Mark N2O tests not applicable when the machine is not configured for N2O.

Expected outcome: Each applicable flow-sensor reading is within the manufacturer-listed range.

12. Perform the Flow-Control and Proportioning Tests

N2O Flow Settings

Record oxygen concentration or measured O2 flow.

O2 Flow Settings

Record oxygen concentration or measured N2O flow.

Expected outcome: The proportioning system prevents delivery outside the manufacturer-listed ranges.

13. Complete Ventilator Calibrations and Safety Tests

Perform the applicable procedures:

Expected outcome

14. Test the Auxiliary O2 Flowmeter, If Equipped

Action

Expected outcome: Measured flow is 2.5–3.5 l/min.

15. Test the Integrated Suction Regulator, If Equipped

Record actual readings for each applicable test.

16. Test ACGO and Passive AGSS, If Equipped

ACGO Pressure-Relief Valve

Expected outcome: The pressure-relief weight relieves and pops off.

Passive AGSS Positive-Relief Test

Expected outcome: Pressure rise does not exceed 10 cmH2O.

17. Perform the Flush-Flow Test

Action: Perform the O2 flush test using the manual setup and time the bellows filling.

Expected outcome: Bellows fill in 1.3–3 seconds.

18. Test Vaporizer Back Pressure and Interlock

Action

Expected outcome

This procedure does not replace vaporizer-specific maintenance required by the applicable vaporizer service documentation.

19. Calibrate the Airway Module, If Installed

Use the procedure and certified calibration gas applicable to the installed module.

Manufacturer checklist calibration points:

Expected outcome: The displayed values calibrate to the specified gas concentrations and the calibration completes without error.

20. Perform Alarm Tests

Verify activation of:

Expected outcome: Each alarm activates under the manufacturer test condition and provides the expected visual and audible indication.

21. Perform the Final System Checkout

Run the complete onboard Full Test after all maintenance, parts replacement, inspection, and calibration work is complete.

Expected outcome: Full Test passes with no unresolved failures.

Electrical Safety Testing

The manufacturer checklist includes electrical safety testing as part of planned maintenance.

Ground Resistance

The Carestation 600 Series manufacturer limit is stricter than the generic 0.5 Ω field often used for other equipment.

Ground resistance measured in ohms, manufacturer expected < 0.20 Ω:

Result =

Pass / Fail =

Leakage Current

Record the supply-voltage category used.

Touch/chassis leakage measured in microamps:

Supply voltage = 100/120 Vac or 220/240 Vac

Manufacturer limit used:

Result =

Pass / Fail =

Do not replace these manufacturer checklist values with a generic facility limit unless the facility has formally evaluated and approved that change through its maintenance program.

Final Machine Checks

Before returning the system to clinical use, verify:

Pass / Fail Criteria

Pass

The PM passes when:

Remove From Service

Remove the machine from service when any of the following remain unresolved:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide