GE Healthcare Panda Warmer

Preventive Maintenance Procedure

On this page

Asset Type

Infant Warmer

Manufacturer

GE Healthcare

Model

Panda Warmer

Manual Reference

Panda and Giraffe Warmers Service Manual, M1128921 Revision J, May 2017

Manufacturer Recommendation / AEM Status

Applicability

This procedure applies to the Panda iRes Warmer, Panda Freestanding Warmer, and Panda Wall Mount Warmer unless a step is identified as configuration-specific.

Scale-related requirements apply only to a Panda iRes Warmer equipped with an optional in-bed scale. Resuscitation system maintenance applies only when the applicable system is installed; the manual identifies the optional resuscitation assembly for bedded and freestanding models, not the wall mount model.

Manufacturer-Listed Annual Activities

Perform electrical safety checks as described in service manual section 2.2.3.

If present, calibrate the scale as described in section 3.1. For the Panda family, this applies to a Panda iRes Warmer equipped with the optional in-bed scale.

If a resuscitation system is installed, perform the resuscitation system service checkout as described in the separate Resuscitation Systems Service Manual.

Manufacturer-Listed Two-Year Activities

Replace the lithium-ion battery as described in service manual section 5.4.3.

If the applicable resuscitation/blender system is installed, perform complete blender maintenance as described in the separate Resuscitation Systems Service Manual.

Manufacturer-Listed Parts and Accessories

Scale calibration test weight for equipped Panda iRes models: 5 kg Certified Test Weight, GE part number 6600-0209-800.

Masimo SET 1 SpO2 electrical leakage test cable: 2006036-001.

Masimo SET 2 SpO2 electrical leakage test cable: M1230051.

Nellcor SpO2 electrical leakage test cable: M1237129.

Post-Repair Safety Inspection Requirement

After completing a repair, perform the Checkout Procedure, a final electrical safety check, and a leakage current test. If any checkout step fails, remove the unit from service and repair it before clinical use.

AEM Note

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility's approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

This procedure covers annual electrical safety testing; annual scale calibration and verification for equipped Panda iRes models; annual resuscitation system checkout reference for equipped Panda iRes or Freestanding models; two-year lithium-ion battery replacement; two-year blender maintenance reference when applicable; SpO2 patient lead leakage testing when installed; configuration-specific documentation; and post-repair safety inspection documentation.

Required Tools and Equipment

Approved electrical safety analyzer.

5 kg Certified Test Weight, GE part number 6600-0209-800, for Panda iRes models with a scale installed.

Applicable SpO2 electrical leakage test cable, if SpO2 is installed: Masimo SET 1 cable 2006036-001; Masimo SET 2 cable M1230051; or Nellcor cable M1237129.

Small screwdriver for lithium-ion battery removal when performing the two-year battery replacement.

Tools and test equipment required by the separate Resuscitation Systems Service Manual when applicable.

Step-by-Step PM Procedure

1. Identify Panda Warmer Configuration and PM Scope

Identify the warmer as a Panda iRes Warmer, Panda Freestanding Warmer, or Panda Wall Mount Warmer.

Determine whether the unit is equipped with a scale, SpO2 module, resuscitation system, or blender requiring separate maintenance, and whether lithium-ion battery replacement is due at this service interval.

Inspect for obvious damage, missing components, damaged power cord, contamination, or any condition that would make continued testing or clinical use inappropriate.

Expected outcome: Correct configuration-specific PM activities are identified before testing begins.

2. Perform Annual Electrical Safety Testing

Using an approved electrical safety analyzer, perform ground continuity testing first, followed by leakage testing. Wait at least 30 seconds after power-on before taking leakage readings.

Ground Resistance Check

Measure between the ground pin of the line cord plug and exposed metal of the electronic enclosure.

Manufacturer limit: less than 0.2 Ω.

Result = ______ Ω
Pass / Fail = ______

Measure between the ground pin of the line cord plug and the ground pin of the appliance outlet connectors.

Manufacturer limit: less than 0.1 Ω.

Result = ______ Ω
Pass / Fail = ______

Important: Do not substitute a generic ground resistance limit for this procedure. The manual provides manufacturer-specific limits of less than 0.2 Ω and less than 0.1 Ω for the identified measurement points.

Earth / Chassis Leakage Current — 115 VAC Rated Device

Normal condition, all possible operating modes: manufacturer limit less than 180 µA.

Result = ______ µA
Pass / Fail = ______

Single-fault condition, if required by local ordinance or facility policy: manufacturer limit less than 200 µA.

Result = ______ µA
Pass / Fail = ______

Earth / Chassis Leakage Current — 220/230/240 VAC Rated Device

Use these limits only for applicable voltage-rated devices.

Normal condition, all possible operating modes: manufacturer limit less than 360 µA.

Result = ______ µA
Pass / Fail = ______

Single-fault condition, if required by local ordinance or facility policy: manufacturer limit less than 500 µA.

Result = ______ µA
Pass / Fail = ______

Expected outcome: All applicable electrical safety measurements meet manufacturer-listed limits.

3. Perform SpO2 Patient Lead Leakage Testing, If Installed

If the Panda Warmer is equipped with an SpO2 module, use the applicable manufacturer-listed leakage test cable. Measure leakage between all SpO2 connector leads shorted together and ground after the required 30-second power-on stabilization period.

AC Patient Lead Leakage

Normal condition, all possible operating modes: manufacturer limit less than 100 µA.

Result = ______ µA
Pass / Fail = ______

Single-fault condition, if required by local ordinance or facility policy: manufacturer limit less than 500 µA.

Result = ______ µA
Pass / Fail = ______

DC Patient Lead Leakage

Normal condition, all possible operating modes: manufacturer limit less than 10 µA.

Result = ______ µA
Pass / Fail = ______

Single-fault condition, if required by local ordinance or facility policy: manufacturer limit less than 50 µA.

Result = ______ µA
Pass / Fail = ______

Expected outcome: SpO2 patient lead leakage measurements pass manufacturer-listed criteria when the option is installed.

4. Calibrate and Verify the Scale, Panda iRes With Scale Only

Perform this step only on a Panda iRes Warmer equipped with the optional in-bed scale. Do not apply this step to Panda Freestanding or Panda Wall Mount units.

Use a 5 kg Certified Test Weight, GE part number 6600-0209-800.

Clear objects from the bed so only the mattress and clear plate remain on the scale. Turn the warmer off and ensure the bed is level. Enter Service Mode by powering on while holding the Help and Alarm Silence keys until two tones are heard. Select Scale Calibration, then Manual Calibration. Enter the gram value of the certified test weight. Follow the on-screen tare prompt. Apply the test weight to the center of the bed when prompted. Save the calibration if successful. Power cycle the unit. Verify scale accuracy using the certified test weight.

Manufacturer Acceptance Criterion

Displayed weight must equal the test weight within ±10 g.

Result = ______ g
Pass / Fail = ______

Required Scale Checkout Following Calibration

Center weight check with 5 kg weight: calculated value must be 5000 ±10.

Four off-center checks with 5 kg weight: each calculated value must be 5000 ±10.

Load cell stability check: displayed SC values must not change by more than 5 counts total.

Center result = ______
Corner B1 result = ______
Corner B2 result = ______
Corner B3 result = ______
Corner B4 result = ______
Load cell stability result = ______ counts
Pass / Fail = ______

Expected outcome: Scale calibration and associated verification procedures pass for an equipped Panda iRes Warmer.

5. Perform Resuscitation System Service Checkout, If Equipped

For an equipped Panda iRes Warmer or Panda Freestanding Warmer, perform the annual resuscitation system service checkout according to the applicable Resuscitation Systems Service Manual.

Expected outcome: Resuscitation system service checkout is completed and documented according to its applicable manual.

Not applicable to: Panda Wall Mount Warmer, based on the warmer manual identifying the optional resuscitation assembly for bedded and freestanding configurations only.

Important: The provided manual does not contain the separate resuscitation system checkout procedure or its pass/fail values.

6. Replace the Lithium-Ion Battery Every Two Years

When due, replace the lithium-ion battery as described in the service manual.

Disconnect power before accessing internal assemblies. For a Panda Wall Mount Warmer, remove the heater head from the wall bracket as instructed by the manual before accessing the enclosure. Open the electronic enclosure rear panel. Tilt the power board panel down to access the battery. Note battery polarity before removal. Use a small screwdriver to slide the old lithium-ion battery out of its holder. Install the new battery with the same polarity. Reassemble the unit.

Expected outcome: New lithium-ion battery is installed with correct polarity and the warmer is properly reassembled.

Why it matters: The lithium-ion battery supplies the power failure alarm and supports memory circuits during a power failure.

7. Perform Two-Year Blender Maintenance, If Applicable

For Panda iRes or Panda Freestanding units equipped with the applicable blender/resuscitation configuration, perform complete blender maintenance according to the separate Resuscitation Systems Service Manual.

Expected outcome: Blender maintenance is completed and documented under the applicable separate manual.

Not applicable to: Panda Wall Mount models unless separate manufacturer documentation establishes otherwise.

8. Complete Post-Repair Safety Inspection When Applicable

After repairs, including lithium-ion battery replacement performed under the repair procedure chapter, perform the applicable Checkout Procedure, final electrical safety testing, and leakage current testing.

Battery Replacement Verification

Record the current control mode and temperature settings. Wait one minute. Unplug the warmer from the wall outlet. Verify the power failure alarm sounds and the power failure indicator illuminates. Wait one to two minutes. Reconnect the warmer to the wall outlet. Verify the alarm cancels and the warmer returns to the previously displayed mode and temperature settings.

Expected outcome: The unit passes checkout, electrical safety, leakage current, and applicable power failure alarm/memory verification.

Note: If the power failure alarm is tested for the full 10-minute minimum operation period, operate the warmer for at least 2 hours to recharge the battery before patient use. The manual lists total recharge time as 10 hours.

Electrical Safety Testing

Ground Resistance

Ground resistance measured in ohms. Manufacturer-specific criteria apply to this device:

Ground resistance from line cord ground pin to exposed metal of the electronic enclosure, expected less than 0.2 Ω.

Result = ______ Ω
Pass / Fail = ______

Ground resistance from line cord ground pin to appliance outlet ground pin, expected less than 0.1 Ω.

Result = ______ Ω
Pass / Fail = ______

Touch / Chassis Leakage — 115 VAC Unit

Touch/chassis leakage measured in microamps:

Facility limit used = Manufacturer limit less than 180 µA, normal condition, all operating modes
Result = ______ µA
Pass / Fail = ______

Facility limit used = Manufacturer limit less than 200 µA, single-fault condition if required locally
Result = ______ µA
Pass / Fail = ______

Touch / Chassis Leakage — 220/230/240 VAC Unit, If Applicable

Touch/chassis leakage measured in microamps:

Facility limit used = Manufacturer limit less than 360 µA, normal condition, all operating modes
Result = ______ µA
Pass / Fail = ______

Facility limit used = Manufacturer limit less than 500 µA, single-fault condition if required locally
Result = ______ µA
Pass / Fail = ______

SpO2 Patient Lead Leakage, If Installed

AC patient lead leakage, normal condition, manufacturer limit less than 100 µA.

Result = ______ µA
Pass / Fail = ______

DC patient lead leakage, normal condition, manufacturer limit less than 10 µA.

Result = ______ µA
Pass / Fail = ______

Single-fault testing completed if required by facility policy or local ordinance: Yes / No / Not Applicable = ______

Pass / Fail Criteria

Pass

Correct Panda configuration is documented. Annual electrical safety testing passes manufacturer-listed limits. Scale calibration and checkout pass for equipped Panda iRes models. SpO2 patient lead leakage testing passes when installed. Resuscitation system service checkout is completed according to the applicable separate service manual when equipped. Two-year battery and blender activities are completed when due. No unresolved unsafe condition remains.

Remove From Service

Remove the unit from service if electrical safety measurements exceed manufacturer-listed limits; Panda iRes scale calibration or required checkout fails; SpO2 patient lead leakage exceeds manufacturer-listed limits; required resuscitation or blender maintenance fails or cannot be completed; post-battery-replacement power failure alarm or memory restoration testing fails; any required post-repair checkout step fails; or the unit has damaged, missing, worn, distorted, or contaminated components affecting safe operation.

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.