Getinge Servo-i

Preventive Maintenance Procedure

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-i

Manual Reference

Servo-i Ventilator System V3.0 User’s Manual, English edition, Order No. 65 14 892 E313E

Manufacturer Recommendation / AEM Status

The provided user manual states that the Servo-i must receive regular service by professionals with specialized training and that all service must be recorded in a service log. It does not specify a calendar-based preventive maintenance interval, detailed service checklist, electrical-safety limits, or scheduled replacement interval for the cooling-fan filter.

Manufacturer-Listed Items Supported by This Manual

Manufacturer-Listed Frequencies and Intervals

Battery modules displaying Replace battery must be replaced immediately. A module displaying Expires soon should be scheduled for replacement even when the status display still indicates usable operating time.

AEM Note

This procedure uses the manufacturer-listed operational checks available in the user manual but is not a substitute for the Getinge or MAQUET Servo-i service manual, service-card procedures, or official preventive-maintenance checklist.

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

The manual specifically identifies the MAQUET blue test tube for the internal Pre-Use Check. It does not provide a complete PM test-equipment list.

Step-by-Step PM Procedure

1. Remove the Ventilator From Clinical Use

Confirm that the Servo-i is not connected to a patient.

Expected outcome: The ventilator is available for controlled inspection and testing without affecting patient care.

2. Review Service History and Configuration

Review:

Identify whether the unit is configured as Servo-i Infant, Adult, or Universal.

Expected outcome: Known recurring conditions and configuration-dependent requirements are identified before testing.

3. Perform an External Visual Inspection

Inspect the complete system for:

Unused module compartments should contain the appropriate empty module to protect the electrical contacts from dust and fluid exposure.

Expected outcome: The ventilator is structurally complete, stable, clean, and free of conditions that could affect safe operation.

4. Inspect the Power Cord and AC Inlet

Inspect:

Connect the ventilator to a known-good grounded AC outlet and confirm that the green AC power indicator operates.

Expected outcome: The power cord and inlet are undamaged, and the ventilator consistently recognizes AC power.

5. Inspect Air and Oxygen Hoses

Inspect both gas hoses and fittings for:

Connect known-good medical air and oxygen supplies.

Expected outcome: Both gas supplies are connected correctly and remain stable without visible damage or leakage.

6. Inspect Battery Modules

Confirm:

Access the Battery Status Window and record:

Replace a battery immediately when Replace battery is displayed. Plan replacement when Expires soon appears.

The manual lists battery-module lifetime as 2.5 years from manufacture date.

Expected outcome: At least two serviceable battery modules are recognized and do not display an unacceptable status instruction.

7. Inspect the Expiratory Cassette and Patient Connections

Inspect the expiratory cassette for:

Inspect the inspiratory outlet, expiratory inlet, viral or bacterial filter arrangement, patient circuit, humidifier connections, Y-piece, and optional accessories.

Expected outcome: The expiratory cassette and breathing-system connections are clean, dry, complete, and properly installed.

8. Start the Complete Pre-Use Check

Connect:

Power on the Servo-i and select Yes when prompted to begin the automatic Pre-Use Check. Follow all on-screen instructions.

Expected outcome: The automatic test sequence begins without technical-error alarms.

9. Complete the Internal Test Connection

Connect the MAQUET blue test tube between the inspiratory outlet and expiratory inlet when instructed. Do not substitute an unapproved test tube.

The automated sequence evaluates:

Expected outcome: Each internal test reports Passed.

Manufacturer-listed internal leakage criterion:
Allowed internal leakage: ≤ 10 mL/min at 80 cmH₂O
Result =

10. Verify AC-to-Battery Transfer

When instructed:

The manual indicates that the connected batteries must provide at least ten minutes of total remaining operation for the battery-switch test to pass.

Expected outcome: Ventilation power transfers between AC and batteries without interruption or unexpected shutdown.

11. Connect and Test the Patient Breathing System

Connect the complete breathing system, including installed accessories that will normally be used.

When instructed:

When patient tubing is replaced, repeat the circuit test and compliance calculation.

Expected outcome: Patient circuit test passes and circuit compliance is measured successfully.

Manufacturer-listed patient-circuit leakage criterion:
Allowed patient-circuit leakage: ≤ 80 mL/min at 50 cmH₂O
Result =

12. Verify Pressure and Safety-Valve Testing

Confirm that the automated test successfully calibrates and checks the inspiratory and expiratory pressure transducers. Confirm that the safety-valve test passes.

Manufacturer-listed safety-valve setting:
Safety-valve opening pressure: 117 ± 3 cmH₂O during the Pre-Use Check adjustment or test
Result =

Expected outcome: Pressure-transducer and safety-valve tests report Passed.

13. Verify O₂ Cell or O₂ Sensor Test

Confirm that the automated test successfully evaluates the oxygen measurement system at 21% and 100% oxygen. Both air and oxygen must be available for this test.

Record:

An independent oxygen analyzer check may be added under facility policy, but the provided manual does not give a PM accuracy-test procedure or tolerance for comparison with an external analyzer.

Expected outcome: The O₂ cell or sensor passes calibration without instability or replacement messages.

14. Verify Flow Measurement and Expiratory Cassette Operation

Confirm that the Pre-Use Check:

Expected outcome: Inspiratory and expiratory flow tests report Passed.

15. Verify Audible and Visual Alarms

Confirm:

Generate appropriate non-clinical alarm conditions using the approved facility test setup.

Expected outcome: Audible and visual alarm indications are clear, consistent, and appropriate to the generated condition.

16. Test the External Alarm Output When Installed

When the Pre-Use Check displays the external alarm-output test:

The manual describes the external alarm as a nonguaranteed alarm connection and recommends establishing a Pre-Use Check routine for the application.

External alarm output installed = Yes / No
Test result =

Expected outcome: The external system receives the intended alarm output.

17. Check User Interface and Controls

Verify operation of:

Confirm that values can be selected, adjusted, confirmed, and canceled normally.

Expected outcome: All accessible controls respond consistently without dead areas, sticking, or unintended operation.

18. Complete Functional Ventilation Check

Connect an approved test lung.

Verify:

Do not independently alter clinical configuration values beyond the approved test setup.

Expected outcome: The Servo-i ventilates the test lung consistently and displays stable measured values without unresolved alarms.

19. Confirm Final Pre-Use Check Status

Complete the test and press OK to log the Pre-Use Check.

Document the final status:

A Failed result requires correction and repetition of the test. Do not return the unit to service with a persistent malfunction.

Expected outcome: Final Pre-Use Check status is Passed.

Electrical Safety Testing

The provided user manual classifies the Servo-i as Class I, Type B equipment and states that the power cord must be connected to a properly grounded AC outlet. It does not provide ground-resistance or leakage-current PM limits.

Complete electrical safety testing according to facility policy, the applicable Getinge service documentation, and the safety standard adopted by the facility.

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:
Result =

Touch/chassis leakage measured in microamps:
Facility limit used =
Result =
Pass / Fail =

Do not claim a manufacturer leakage-current limit from this user manual.

Pass / Fail Criteria

PM Pass Criteria

Remove From Service Criteria

Remove the Servo-i from service when:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide