Hamilton C3

Preventive Maintenance Procedure

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

C3

Manual Reference

Hamilton Medical HAMILTON-C3 Operator's Manual, software version 2.0.x, USA document 10103184.00. Confirm the installed software, adult/pediatric or neonatal capability, high- or low-pressure oxygen configuration, battery count, galvanic O2 cell, CO2/SpO2 options, communication interfaces, trolley, humidifier, nebulizer, and accessories before service.

Manufacturer Recommendation / AEM Status

Hamilton assigns service-related annual PM and the CO2 accuracy check to authorized service personnel using the controlled service manual. This page separates operator-accessible inspection and preoperational testing from restricted service work. Any facility change to the manufacturer schedule belongs under its approved AEM program, documented risk assessment, service history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Safety: Never perform service or maintenance while a patient is connected. Provide alternate ventilation, place the unit in standby, power it down when required, and reprocess applicable components. Refer internal work, annual service-related PM, and CO2 accuracy testing to properly trained and Hamilton-authorized personnel.

Step-by-Step PM Procedure

1. Remove the Ventilator From Clinical Use

Action: Confirm no patient is connected, provide alternate ventilation, place the C3 in standby, shut it down safely when required, label it unavailable, and clean or disinfect the ventilator and reusable patient-path components according to their instructions.

Expected outcome: Maintenance can proceed without interrupting therapy or exposing personnel to contamination.

2. Verify Identity, Configuration, and Service History

Action: Record asset tag, serial number, software revision, operating hours from System > Info, patient-group capability, oxygen configuration, battery count and age, O2-cell status, CO2/SpO2 options, interfaces, trolley, and accessories. Review technical alarms, repairs, field actions, battery history, filter changes, and the last authorized annual service.

Expected outcome: The unit matches its CMMS record and all applicable tasks, options, parts, and authorization boundaries are identified.

3. Perform the Physical Inspection

Action: Inspect the enclosure, handle, display, touchscreen, control knob, alarm lamp, speaker openings, ventilation outlet, pressure-line and flow-sensor connections, expiratory-valve area, oxygen inlets, cooling openings, communication connectors, AC/DC supplies, cords, plugs, strain reliefs, batteries, trolley, rails, humidifier mount, brackets, hardware, and accessories. Confirm labels are legible and check for cracks, contamination, corrosion, fluid entry, loose parts, damaged pins, exposed conductors, blocked airflow, or insecure mounting.

Expected outcome: The complete system is clean, secure, legible, undamaged, and free of visible electrical, pneumatic, mechanical, or mounting hazards.

4. Inspect the Monthly Filters

Action: Remove and inspect the rear-panel fan filters and the white air-intake filters outside the HEPA filter. Check for dust, lint, deformation, moisture, or damage. Replace as needed, and never reinstall wet or damaged filter media.

Expected outcome: Clean, dry, undamaged filters are correctly seated with unrestricted airflow, and the monthly inspection is documented.

5. Replace the Annual HEPA Filter

Action: Verify the correct air-intake HEPA filter, remove the old filter according to the operator manual, inspect its housing and seals, install the new filter in the proper orientation, and secure the cover.

Expected outcome: A correct new HEPA filter is fully seated, sealed, and documented at the yearly interval.

6. Inspect and Recharge the Batteries

Action: Inspect each battery and compartment for damage, swelling, contamination, corrosion, or poor retention. Review age, runtime complaints, and history. At the six-month interval, connect the ventilator to primary power for at least four hours. Verify battery recognition, charge indication, estimated runtime when available, and operation on each installed battery.

Expected outcome: Each battery is intact, securely retained, fully charged, recognized, and able to support normal power transfer without a battery fault.

7. Evaluate the Oxygen Cell and Gas Path

Action: Inspect the galvanic O2 cell, oxygen inlets, hoses, fittings, low-pressure reservoir or high-pressure connection, and related accessories. Replace the O2 cell when depleted. Connect the correct oxygen source and confirm secure, leak-free connections.

Expected outcome: Gas-path components are correct and intact, and the O2 cell is recognized and ready for calibration.

8. Complete Authorized Annual Service

Action: Have Hamilton-authorized service personnel perform the yearly service-related preventive maintenance using the software- and serial-specific service manual. When CO2 is installed, include the required accuracy check. Record the service report, parts, measurements, adjustments, software actions, and post-service verification.

Expected outcome: All restricted annual tasks are completed with no unresolved service message, fault, or overdue component.

9. Run the Startup Self-Test

Action: Connect primary power and the correct oxygen source, assemble the specified test circuit, and turn on the ventilator. Observe the alarm lamp and audible signal during startup, then confirm the C3 reaches standby without an unresolved technical alarm.

Expected outcome: Startup self-test, alarm indicators, display, controls, and standby transition complete normally.

10. Perform the Circuit Tightness Test

Action: From Standby, open the preoperational check and run the Tightness test with the complete intended circuit. Follow the on-screen instructions to occlude and release the patient end. Repeat after changing the circuit or a pneumatic component.

Expected outcome: The ventilator reports a passing tightness result and stores a current completion date and time.

11. Calibrate the Flow Sensor and Applicable Sensors

Action: Calibrate the correct adult/pediatric flow sensor, neonatal flow sensor, or neonatal circuit for the selected patient group and mode. Reverse the flow sensor and use the calibration adapter when prompted. Perform O2-cell calibration and zero the installed CO2 sensor and airway adapter when required.

Expected outcome: Every applicable calibration succeeds, with no patient-group, sensor, or circuit mismatch.

12. Verify Power Transfer, Charging, and Alarms

Action: With the C3 operating on primary power, disconnect that source and verify the loss-of-external-power alarm and uninterrupted battery operation. Reconnect power and confirm alarm reset and charging. Generate representative high/low pressure, disconnection, oxygen, apnea, power, and battery alarms using the controlled procedure, then verify visual indication, audible signal, message, priority, silence/reset, and recovery.

Expected outcome: Power transfers without shutdown, batteries charge normally, and every required alarm is detected, announced, displayed, and cleared correctly.

13. Verify Ventilation and Installed Options

Action: Connect a compatible test lung and use the software/configuration-specific procedure to check representative modes, rate, pressure, PEEP, delivered volume, oxygen monitoring, trigger response, and sensor readings. Verify CO2, SpO2, nebulizer, humidifier, communication, nurse-call, and trolley power functions when installed.

Expected outcome: Delivered and measured parameters and all installed options meet the applicable Hamilton criteria with no unresolved alarm or calibration error.

14. Perform Electrical Safety Testing

Action: Apply the current Hamilton service procedure and facility policy for the exact C3 mains-power, device-class, trolley, and installed applied-part configuration. Test powered accessories as applicable and record the analyzer, method, limits, and measurements.

Expected outcome: Every required electrical-safety measurement passes the documented criterion.

15. Restore the System and Make the Return-to-Service Decision

Action: Remove test equipment, reinstall filters, covers, batteries, and approved accessories, clean surfaces, and configure the intended circuit. Repeat the preoperational check after work affecting the circuit or pneumatic path. Confirm date/time, configuration, charge state, trolley security, and absence of technical alarms. Document parts, measurements, service reports, failures, and corrective work.

Expected outcome: The Hamilton C3 is complete, clean, current on maintenance, and either formally returned to service or clearly quarantined.

Electrical Safety Testing

The public operator manual assigns annual service-related PM to Hamilton-authorized personnel and does not publish one universal electrical-safety limit for every C3 power, trolley, and option configuration. Use the current controlled Hamilton service procedure, the facility's approved analyzer method, and the exact device class and applied parts. Do not copy limits from a C2, T1, or differently configured C3.

Analyzer / serial / calibration due:
Result =

Power configuration, trolley, equipment class, and applied parts:
Result =

Test standard, method, governing limits, and measurements:
Results =
Pass / Fail =

Pass / Fail Criteria

Pass

Remove From Service

Copy/Paste CMMS Work Order Procedure

Do not return the Hamilton C3 to service with a failed preoperational check, circuit leak, unsuccessful calibration, failed alarm or power transfer, unacceptable battery, overdue authorized service, failed electrical-safety result, or any unresolved condition that could interrupt or impair ventilation.