Hamilton G5

Preventive Maintenance Procedure

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

G5

Manual Reference

HAMILTON-G5 Ventilator Service Manual including maintenance, repairs and tests, PN 624093/01, March 2008. Manual coverage is listed as software versions 1.0 and 1.1x. Confirm that this manual revision applies to the device/software in service before using it as the active PM reference.

Manufacturer Recommendation / AEM Status

Manufacturer-Listed PM Frequency

The manufacturer lists complete preventive maintenance and testing at the following interval:

The complete PM/testing workflow includes hospital preventive maintenance review, engineer preventive maintenance, Test Mode procedures, required part replacement, cover installation, electrical safety testing, and completion of operator-manual tests and calibrations.

Manufacturer-Listed Maintenance Items and Intervals

Manufacturer-Listed Parts Referenced for PM

Electrical Safety and Post-Repair Requirements

The complete manufacturer-listed PM workflow includes electrical safety testing.

The manual specifically states that electrical safety testing must be performed after replacement of the Power Management Supply or removal of any ground/earth contact from the ventilator.

The manual also directs completion of the full test sequence after repair or adjustment and requires electrical safety testing and the pre-operational check before returning the ventilator to normal use.

AEM Note

The manufacturer-listed PM interval is annual or every 5,000 operating hours, whichever comes first, with additional maintenance intervals for specific components. A facility performing this PM at a different interval or changing the listed activities should do so under its approved AEM program, risk assessment, maintenance history and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Required or Manual-Supported Tools

Test Mode Equipment Listed in the Manual

Additional Equipment When CO2 Monitoring Is Installed

Step-by-Step PM Procedure

1. Confirm PM Due Status and Device Configuration

Action: Confirm the device identity, software level, operating hours, installed options, and whether the CO2 module or Heliox option is installed. Determine whether the PM is scheduled by interval, triggered by component replacement, or being completed after repair.

Expected outcome: The correct PM scope is identified before testing begins.

Important: If the Heliox option is installed, disconnect the Heliox gas supply and connect air supply gas before performing maintenance and testing procedures.

2. Remove the Ventilator From Clinical Use

Action: Remove the ventilator from patient care before beginning maintenance or Test Mode procedures. Follow facility cleaning and infection-control processes before handling patient-contact accessories.

Expected outcome: The ventilator is available for complete testing without interrupting patient ventilation.

3. Review Hospital Preventive Maintenance Items

Action: Confirm that hospital-maintenance activities are being performed at the manufacturer-listed intervals, including patient circuit/accessory replacement or reprocessing, pre-operational checks, gas inlet water trap inspection, breathing circuit and water trap inspection, and fan filter inspection.

Expected outcome: Routine user/hospital-maintenance items are complete and any deficiencies are corrected or documented.

4. Inspect Gas Inlet Microfilters

Action: Inspect the gas inlet microfilter elements.

Expected outcome: Microfilter elements are clean and suitable for continued service.

Pass / corrective action: Replace the gas inlet microfilter elements if discolored. Manufacturer-listed microfilter element part number: 155714.

5. Inspect Scheduled Replacement Items

Action: Review operating hours, age and service history for interval-based replacements.

Expected outcome: Replacement items due by time or operating hours are identified and addressed.

6. Enter Test Mode and Perform the Complete Test Sequence

Action: Connect the ventilator to mains power, switch it on, enter Test Mode and perform the manufacturer-listed test units in order. Complete the Hamilton G5 Ventilator Test Report during testing.

Expected outcome: All required Test Mode checks pass or any failed test is corrected before proceeding.

Manufacturer-listed Test Mode sequence:

Important: If testing identifies a part that must be replaced, replace the part and then repeat the complete series of tests.

7. Record Key Test Mode Results

Action: Document required test values using the Hamilton test report or facility-equivalent documentation.

Expected outcome: Test results demonstrate that critical pneumatic, electronic, battery and alarm functions remain within manufacturer-listed criteria.

8. Perform O2 Cell Calibration and Checks

Action: Verify an oxygen cell is installed, connect air and oxygen supplies, perform the O2 cell calibration and complete the listed checks.

Expected outcome: Oxygen cell calibration completes and displayed oxygen measurements remain within listed limits.

9. Perform Mixer and Pneumatic Safety Checks

Action: Perform tank tightness, mixer valve leakage, mixer gain, tank overpressure and associated pneumatic checks.

Expected outcome: Mixer and pressure-limiting systems meet manufacturer-listed criteria.

10. Perform Flow Sensor, Valve and Nebulizer Checks

Action: Complete flow sensor circuit, ambient valve, inspiratory valve, expiratory valve and nebulizer testing according to the manual.

Expected outcome: Gas delivery, pressure control, exhalation control, rinse-flow function and nebulizer output remain within manufacturer-listed criteria.

11. Perform CO2 Sensor Accuracy Check, If Installed

Action: When the HAMILTON MEDICAL CAPNOSTAT 5 CO2 monitoring option is installed, perform the CO2 sensor accuracy check using the manufacturer-listed setup and calibration gas.

Expected outcome: CO2 measurement meets the acceptable PetCO2 range for the measured ambient temperature and barometric pressure.

12. Secure Adjustments and Reassemble the Ventilator

Action: Lock adjustment potentiometers using proprietary locking paint or nail varnish when adjustments were made. Close and secure all ventilator covers. Verify all removed accessories, cables, tubing and hardware are correctly restored.

Expected outcome: The ventilator is fully assembled and ready for safety and operational verification.

13. Perform Electrical Safety Testing

Action: Perform the electrical safety checks listed below using appropriate test equipment.

Expected outcome: Grounding, leakage-current and insulation-resistance results meet the manufacturer-listed IEC 60601-1 criteria for the installed configuration.

14. Complete Final Operational Verification

Action: Exit Test Mode. Switch the ventilator off and back on. Complete the pre-operational check described in the Hamilton G5 Operator's Manual. Complete any additional tests and calibrations required by the operator's manual. Confirm that unresolved technical faults are not present.

Expected outcome: The ventilator successfully completes final verification and is suitable for disposition according to facility policy.

Electrical Safety Testing

Manufacturer-Listed Test Classification

Important: Type BF testing with the CO2 module installed requires the CO2 module electrical safety test cable, PN 159171.

Visual Grounding Checks

Action: Inspect the yellow/green internal ground wires near the mixer block. Pull each ground wire to verify tightness. Check that contact strips in the main housing are secure.

Result: Pass / Fail =

Internal Ground Continuity Checks

Protective Earth Resistance

Protective earth resistance measured in ohms, including power cord, manufacturer-listed expected limit ≤ 0.2 Ω:

Result =

Pass / Fail =

Type B Testing — CO2 Module Not Installed

Type BF Testing — CO2 Module Installed

Pass / Fail Criteria

PM Pass Criteria

Pass the PM when:

Remove From Service Criteria

Remove the ventilator from service when:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide