Hamilton T1

Preventive Maintenance Procedure

On this page

Asset Type

Ventilator

Manufacturer

Hamilton

Model

T1

Manual Reference

Hamilton Medical HAMILTON-T1 Operator's Manual, software 3.0.x, document 10101879.01. Confirm the installed software, adult/pediatric or neonatal configuration, high- or low-pressure oxygen setup, battery count, galvanic or paramagnetic O2 sensor, CO2/SpO2 options, communication interfaces, trolley or transport mounting, humidifier, and accessories before service.

Manufacturer Recommendation / AEM Status

Hamilton assigns annual service-related PM, battery service, and CO2 accuracy verification to authorized service personnel using the controlled service manual. This page separates operator-accessible inspection and testing from restricted service work. Any facility change to the manufacturer schedule belongs under its approved AEM program, documented risk assessment, service history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Safety: Never perform maintenance with a patient connected. Provide another means of ventilation, place the unit in standby, power it down when required, and reprocess applicable components. Refer internal, annual service, battery service, and CO2 accuracy work to properly trained and authorized personnel.

Step-by-Step PM Procedure

1. Remove the Ventilator From Clinical Use

Action: Confirm no patient is connected, provide alternate ventilation, place the T1 in standby, shut it down safely when required, label it unavailable, and clean or disinfect the ventilator and reusable patient-path components.

Expected outcome: Maintenance can proceed without interrupting therapy or exposing personnel to contamination.

2. Verify Identity, Configuration, and Service History

Action: Record asset tag, serial number, software revision, operating hours from System > Info, patient-group capability, oxygen configuration, battery count and age, O2 sensor type, CO2/SpO2 options, interfaces, mount, and accessories. Review faults, repairs, field actions, battery history, filter changes, and last authorized annual service.

Expected outcome: The unit matches its CMMS record and all applicable tasks, options, parts, and service boundaries are identified.

3. Perform the Physical Inspection

Action: Inspect the enclosure, handle, display, touchscreen, rotary control, alarm lamp, speaker openings, ports, circuit and pressure-line connections, expiratory-valve area, oxygen inlets, cooling openings, USB/communication connectors, AC/DC power supplies, cords, plugs, strain reliefs, batteries, transport mount or trolley, rails, brackets, hardware, and accessories. Confirm labels are legible and check for cracks, contamination, corrosion, fluid entry, loose parts, damaged pins, exposed conductors, or blocked airflow.

Expected outcome: The complete transport system is clean, secure, legible, undamaged, and free of visible electrical, pneumatic, mechanical, or mounting hazards.

4. Inspect the Fan and Air-Intake Filters

Action: Remove and inspect the rear-panel fan filters and the white air-intake filters on the outside of the HEPA filter. Check for dust, lint, deformation, moisture, or damage. Replace filters as needed; do not reinstall wet or damaged media.

Expected outcome: Clean, dry, undamaged filters are correctly seated with unrestricted airflow, and the monthly inspection is documented.

5. Replace the Annual HEPA Filter

Action: Verify the correct HEPA filter for the installed configuration, remove the old filter according to the operator manual, inspect its housing and seals, install the new filter in the proper orientation, and secure its cover.

Expected outcome: A correct new HEPA filter is fully seated, sealed, and documented at the yearly interval.

6. Review Battery Condition and Complete Due Service

Action: Inspect each battery and its compartment for damage, swelling, contamination, corrosion, or poor retention. Confirm the six-month four-hour recharge has been completed. Record battery age and service history; route batteries for annual authorized service and replace them when the controlled procedure indicates or at the applicable life limit. After service, verify charging indications and operation on each installed battery.

Expected outcome: Batteries are intact, charged, within their approved service status, retained securely, and capable of supporting normal power transfer without a battery alarm.

7. Evaluate the Oxygen Sensor and Gas Path

Action: Identify the installed O2 sensor type. Inspect oxygen inlets, hoses, fittings, low-pressure reservoir or high-pressure connection, and related accessories. Replace a depleted galvanic O2 sensor when indicated; do not schedule replacement of a paramagnetic sensor as though it were galvanic. Connect the correct oxygen source and check for secure, leak-free connections.

Expected outcome: Gas-path components are correct and intact, and the installed O2 sensor is recognized and ready for calibration or functional testing.

8. Complete Authorized Annual Service Tasks

Action: Have Hamilton-authorized service personnel perform the yearly service-related preventive maintenance using the software- and serial-specific service manual. Include battery service and the CO2 accuracy check when those items apply. Record the service report, parts replaced, measurements, adjustments, software actions, and post-service verification.

Expected outcome: All restricted annual tasks are completed by qualified personnel with no unresolved service message, fault, or overdue component.

9. Run the Startup Self-Test

Action: Connect primary power and the correct oxygen supply, assemble the specified test breathing circuit, and turn on the ventilator. Observe the alarm lamp sequence and audible signal during startup, then confirm the unit reaches standby without an unresolved technical alarm.

Expected outcome: The startup self-test, alarm indicators, display, controls, and standby transition complete normally.

10. Perform the Circuit Tightness Test

Action: From Standby, open the preoperational check and run the Tightness test with the complete intended breathing circuit. Follow the on-screen prompts to occlude and release the patient end. Repeat after any circuit or pneumatic component is changed.

Expected outcome: The ventilator reports that the tightness test passes and stores a current completion date and time.

11. Calibrate the Flow Sensor and Applicable Sensors

Action: Calibrate the correct adult/pediatric flow sensor, neonatal flow sensor, or neonatal circuit for the selected patient group and mode. Reverse the flow sensor and use the calibration adapter when prompted. Perform O2 sensor calibration when required and CO2 sensor/adapter zero calibration when installed and indicated.

Expected outcome: Each applicable calibration displays a successful result or checkmark; no patient-group, sensor, or circuit mismatch remains.

12. Verify Power Transfer, Charging, and Alarms

Action: With the ventilator operating on primary power, disconnect that source and verify the loss-of-external-power alarm and uninterrupted battery operation. Reconnect primary power and verify alarm reset and charging. Generate representative test alarms using the operator-manual method and confirm the visual indicator, audible signal, message, priority behavior, silence/reset, and recovery.

Expected outcome: Power transfers without shutdown, batteries charge normally, and each required alarm is detected, announced, displayed, and cleared correctly.

13. Verify Ventilation and Installed Options

Action: Connect a compatible test lung and use the software/configuration-specific controlled procedure to check representative modes, rate, pressure, PEEP, delivered volume, oxygen monitoring, trigger response, and sensor readings. Verify CO2, SpO2, nebulizer, humidifier, communication, nurse-call, and transport-power functions when installed and included in the controlled procedure.

Expected outcome: Delivered and measured parameters and all installed options meet the applicable Hamilton acceptance criteria with no unresolved alarm or calibration error.

14. Perform Electrical Safety Testing

Action: Apply the current Hamilton service procedure and facility policy for the exact T1 mains-power supply, device class, and installed CO2/SpO2 or other applied-part configuration. Test all applicable power supplies and powered accessories, record the analyzer method and limits, and compare results with the controlled criteria.

Expected outcome: Every required measurement passes the documented criterion for the exact configuration.

15. Restore the System and Make the Return-to-Service Decision

Action: Remove test equipment, reinstall filters, covers, batteries, and approved accessories, clean surfaces, and configure the intended circuit. Repeat the preoperational check after work affecting the circuit or pneumatic path. Confirm date/time, configuration, charge state, mounting security, and absence of technical alarms. Document all parts, measurements, service reports, failures, and corrective work.

Expected outcome: The Hamilton T1 is complete, clean, current on maintenance, and either formally returned to service or clearly quarantined.

Electrical Safety Testing

The public operator manual assigns service-related annual PM to authorized service personnel and does not provide one universal electrical-safety limit for every T1 power-supply and option configuration. Use the current controlled Hamilton service procedure, the analyzer method required by the facility, and the exact device class and applied-part configuration. Do not copy limits from a C1, MR1, or a differently configured T1.

Electrical-safety analyzer / serial / calibration due:
Result =

Installed power supply and equipment class:
Result =

Applied parts / CO2 / SpO2 / accessories:
Result =

Test standard, method, and governing limits:
Result =

Protective-earth, equipment/touch, and patient leakage results as applicable:
Results / N/A =
Pass / Fail =

Pass / Fail Criteria

Pass

Remove From Service

Copy/Paste CMMS Work Order Procedure

Do not return the Hamilton T1 to service with a failed preoperational check, circuit leak, unsuccessful calibration, failed alarm or power transfer, unacceptable battery, overdue annual authorized service, failed electrical-safety result, or any unresolved condition that could interrupt or impair ventilation.