Masimo Rad-5

Preventive Maintenance Procedure

On this page

Asset Type

Pulse Oximeter

Manufacturer

Masimo

Model

Rad-5

Manual Reference

Masimo Rad-5/5v Operator's Manual, LAB-5724H, order number 33606, revision date March 10, 2022.

Manufacturer Recommendation / AEM Status

The reviewed manual provides performance verification for use following repairs or during routine maintenance but does not specify a calendar-based preventive-maintenance interval.

If the facility establishes an interval, manage it through its approved AEM program, risk assessment, maintenance history, clinical use, and applicable requirements.

What This PM Procedure Covers

Required Tools and Equipment

Do not use undiluted bleach or an unlisted solution. Do not soak, immerse, irradiate, steam sterilize, autoclave, or use ethylene oxide on the device.

Step-by-Step PM Procedure

1. Confirm Device Identification and Configuration

Action: Verify the asset tag, serial number, manufacturer, model, accessories, sensor and cable type, and CMMS record. Confirm this is a Rad-5, not the alarm-free Rad-5v.

Expected outcome: The device matches its record and Rad-5-specific alarm tests apply.

2. Remove the Device From Clinical Use

Action: Confirm the device is not monitoring a patient. Turn it off and disconnect the sensor, patient cable, and serial cable.

Expected outcome: The Rad-5 is ready for safe maintenance.

3. Perform a Physical Inspection

Action: Inspect the case, displays, LEDs, buttons, speaker openings, connectors, battery cover/latch, battery compartment/contacts, patient cable, sensor, and accessories for cracks, contamination, corrosion, leakage, damaged contacts, loose parts, or liquid intrusion. Do not lift by the patient cable or open the case beyond battery replacement.

Expected outcome: No unsafe, damaged, contaminated, or unreliable condition is present.

4. Clean and Dry the Device

Action: With the device off, patient disconnected, and batteries removed, wipe outer surfaces twice or until free of visible residue using a dampened soft cloth with mild detergent/warm water or a listed agent. Prevent liquid entry and dry thoroughly. Do not use abrasives, instruments, brushes, or rough materials on displays.

Expected outcome: The device is visibly clean, completely dry, and undamaged.

5. Inspect and Install the Batteries

Action: Inspect for leakage, corrosion, or damage. Install four new or verified AA alkaline batteries in the marked orientation. Do not use non-alkaline batteries or cells greater than 1.5 V. Lock the cover securely.

Expected outcome: A matched alkaline set is installed and the compartment is intact.

6. Perform the Power-On Self-Test

Action: With patient, sensor, and serial cables disconnected, press Power and observe the approximately five-second startup.

Expected outcome: All available LEDs illuminate, a brief beep sounds, and normal operation begins without a system-failure message.

7. Perform the Key Press Button Test

Action: Press every button except Power and verify an audible beep or appropriate display change.

Expected outcome: Every tested control is acknowledged consistently.

8. Perform the Rad-5 Alarm Limit Test

Action: Record the original settings. Enter the alarm menu, change the high SpO2 limit two points below its current value, accept it, verify it displays, and restore it. Repeat the change/display/restore test for low SpO2, high pulse rate, and low pulse rate. Confirm all four original limits are restored.

Expected outcome: All four alarm limits change, display, and restore correctly.

9. Test LED Brightness

Action: Use the applicable Rad-5 menu and Up/Down keys to cycle through all four brightness levels. Exit with Mode/Enter or normal timeout.

Expected outcome: All four levels are distinguishable and normal operation resumes.

10. Test With the Masimo SET Tester

Action: Power-cycle the oximeter and connect the Masimo SET Tester. Verify:

Restore the original alarm settings.

Expected outcome: All listed ranges, indicators, alarm behavior, timing, and volume controls pass.

Important limitation: A functional tester cannot assess Rad-5/5v accuracy. Record performance verification, not clinical accuracy calibration.

11. Perform Final Sensor and Operational Verification

Action: Disconnect the tester. Connect an authorized cable and sensor. Verify secure connections, sensor recognition, no unresolved messages, appropriate SpO2 and pulse-rate display during a controlled non-patient check, Signal IQ/pulse-bar response, and audible pulse indication. Do not derive clinical accuracy tolerance from another ordinary oximeter or an uncontrolled human reading.

Expected outcome: The installed system recognizes the sensor and operates normally.

12. Complete the Return-to-Service Review

Action: Confirm every applicable step passes, all settings are restored, and no unresolved complaint remains.

Expected outcome: The Rad-5 is ready to return to service according to facility policy.

Electrical Safety Testing

The Rad-5 uses four AA batteries. The manual identifies it as a defibrillation-proof Type BF device but does not list protective-earth resistance or touch/chassis leakage testing in routine maintenance. Do not add inapplicable mains-powered limits. Evaluate separately powered test equipment under its own documentation and facility policy.

Pass / Fail Criteria

PM Pass Criteria

Remove From Service Criteria

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Sources Used

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide