Medtronic Nellcor N-600x Series

Preventive Maintenance Procedure

On this page

Asset Type

Pulse Oximeter

Manufacturer

Medtronic

Model

Nellcor N-600x Series

Manual Reference

OxiMax N-600x Pulse Oximeter with Alarm Management System Service Manual, document 10030881 Rev. B, December 2008; applicable N-600x Operator's Manual and OxiMax sensor instructions.

Manufacturer Recommendation / AEM Status

The N-600x requires no calibration. At least every 24 months, qualified service should inspect mechanical/functional condition and safety-label legibility, perform electrical safety tests, inspect fuses F1/F2 for 0.5 A, 250 V, and complete service-manual performance verification.

Replace the internal battery at least every two years. Replace a battery stored longer than six months or unable to recharge fully. A new fully charged battery provides at least seven hours only under the manual's stated conditions: no audible alarms, no analog/serial devices, and default brightness.

After visible damage or mechanical shock, perform the performance tests immediately. Except for battery charge/performance testing, the manual states performance tests should be the final operation before return to the caregiver.

Any interval/activity change should be managed under the approved AEM program, risk assessment, maintenance history, and applicable requirements.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Confirm Identification and Configuration

Action: Verify asset tag, serial number, model, software version, institutional defaults, power configuration, battery age, sensors/cables, output and nurse-call use, mounting, and accessories.

Expected outcome: The device matches its record and configuration-dependent tests are identified.

2. Remove the Monitor From Clinical Use

Action: Transfer monitoring if needed. Power off and disconnect patient, sensor, data cables, and AC before cleaning/inspection.

Expected outcome: Maintenance proceeds safely without interrupting patient care.

3. Perform the 24-Month Inspection

Action: Inspect enclosure, LCD/backlight, LEDs, controls, speaker, inlet, cord/plug/earth pin, sensor/data ports, battery compartment, mounts, DOC-10, sensors, and accessories for damage, deterioration, contamination, corrosion, liquid intrusion, heat, or strain. Check safety labels and, through qualified service, confirm F1/F2 are each 0.5 A, 250 V.

Expected outcome: No mechanical, functional, electrical, accessory, label, or fuse defect is present.

4. Clean and Disinfect

Action: Lightly wipe surfaces with a soft cloth dampened with nonabrasive cleaner or 70% alcohol/water. For disinfection under policy, use 10% chlorine bleach/tap water. Follow sensor instructions. Do not spray, pour, or spill liquid on equipment or openings. Dry thoroughly.

Expected outcome: Equipment is clean, dry, and undamaged.

5. Review Battery Age and Condition

Action: Confirm age/history. Replace through qualified service if two years old, stored longer than six months, unable to recharge fully, damaged, or unreliable. Check terminal clips after replacement.

Expected outcome: The two-year requirement is met and connections are secure.

6. Charge and Check the Battery

Action: Connect AC, verify the monitor is off and the charging indicator is lit, then charge at least eight hours off or twelve hours on when full charge is required. With a full battery and monitor operating, disconnect AC; verify the AC indicator turns off, operation continues, and Low Battery is off. Reconnect AC.

Expected outcome: Charging and AC/battery transition are stable. Use the approved battery-performance procedure if runtime certification is required.

7. Perform the Power-On Self-Test

Action: Connect AC with no other cables and power on. Within ten seconds verify all LEDs, pixels, and backlight illuminate; indicators remain lit two seconds; NELLCOR/software version display; a three-beep tone followed by a one-second beep sounds; and normal operation begins.

Expected outcome: POST, display, backlight, indicators, speaker, and startup pass.

8. Verify Defaults, Alarm Ranges, and SatSeconds

Action: Record institutional defaults. Follow the service-manual sequence to verify adult/neonate selection, alarm ranges, selection beeps, and SatSeconds cycling OFF, 10, 25, 50, 100, and OFF. Restore institutional settings.

Expected outcome: Controls/ranges work and no unauthorized default change remains.

9. Verify Alarms, Alarm Silence, and Audio Volume

Action: Create SpO2 and pulse alarms. Verify pleth tracks pulse, pulse tone is audible, values flash, and alarm sounds. Verify a 60-second silence: indicator lit, values flashing, pulse tone audible, alarm returns at approximately 60 seconds. Verify 30/60/90/120-second choices and OFF only if ALLOW OFF is enabled. If OFF is enabled, verify flashing silence indicator and three-beep reminder at approximately three-minute intervals. Cycle alarm volume 1-10 and pulse-tone volume, then restore settings.

Expected outcome: Audible/visual alarms, timing, reminders, and volume controls pass.

10. Perform the Sensor LED-Excitation Check

Action: Connect DOC-10 and MAX-A. With the sensor open, verify bright LED. Slowly close it and verify intensity decreases; reopen and verify it increases.

Expected outcome: LED intensity modulation responds normally.

11. Perform SRC-MAX Functional Tests

Action: Connect SRC-MAX through DOC-10 and complete all four tests in sequence:

Expected outcome: Exact ranges and expected waveform/alarm behavior pass.

12. Verify Nurse Call and Analog Outputs When Used

Action: Follow the service-manual setup:

Restore normal output configuration.

Expected outcome: Configured outputs meet listed values.

13. Perform Electrical Safety Testing

Action: Test as Class I Type BF using applicable standards. Connect applied-part leakage through DOC-10 using the shorted 9-pin connector.

Use the exact applicable standard, region, manual edition, and facility criteria.

Expected outcome: Every required test passes.

14. Complete the Return-to-Service Review

Action: Confirm inspection, labels/fuses, cleaning, battery, POST, settings, alarms/audio, sensor excitation, SRC-MAX, outputs, electrical safety, and restoration all pass with no unresolved fault.

Expected outcome: The N-600x is ready to return to service.

Electrical Safety Testing

Electrical safety is required at least every two years and after applicable repair because the N-600x is Class I Type BF mains-powered equipment. Record exact analyzer results, configuration, standard, and limits.

Pass / Fail Criteria

PM Pass Criteria

Remove From Service Criteria

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Sources Used

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide