Mindray WATO EX Series

Preventive Maintenance Procedure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Mindray

Model

WATO EX Series

Manual Reference

WATO EX-55/65 Anesthesia Machine Service Manual, Revision 1.1, March 2009.

Manufacturer Recommendation / AEM Status

Mindray lists a 12-month maintenance interval based on typical use of approximately 2,000 operating hours per year. Equipment used more heavily should be serviced more frequently.

Manufacturer-Listed Annual Maintenance

Manufacturer-Listed Three-Year Replacement Items

After servicing the equipment or replacing components, fully reassemble the machine, complete all necessary calibrations, and perform the applicable checkout tests in the service manual before returning the machine to service.

If the facility performs this PM at a different interval or changes the listed activities, those changes should be managed under the facility’s approved AEM program, documented risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Facility policy may require additional equipment such as an independent oxygen analyzer, vaporizer analyzer, gas-pressure gauges, or documentation of test-equipment calibration status.

Step-by-Step PM Procedure

1. Remove the Unit From Clinical Service

Action:

Expected outcome: The machine is safe to inspect and test without exposing a patient or technician to avoidable hazards.

2. Review Configuration and Maintenance History

Action: Record:

Expected outcome: Testing is matched to the machine’s actual configuration.

3. Replace Manufacturer-Listed Annual Components

The one-year service package is listed as P/N 0621-30-78479.

Replace the applicable components according to the unit’s configuration and service history:

The manual repeatedly uses the wording “as required” while also listing these components under the 12-month replacement schedule. Document each listed component not replaced and the approved reason.

Expected outcome: Applicable annual wear components and seals are renewed and correctly installed.

4. Perform the System Inspection

Action: Confirm that:

Expected outcome: The machine is complete, stable, correctly assembled, and free of visible safety defects.

5. Test Pipeline Gas Supplies

Action:

Expected outcome: Gauges zero correctly, gas-specific connections are correct, pipeline pressures are within range, and low-supply alarms operate.

6. Test Cylinder Supplies

Skip this step when cylinders are not configured.

Cylinder Pressure

Action:

Expected outcome: Each installed reserve cylinder contains sufficient gas.

High-Pressure Leak Test

Action:

Pass criteria:

Replace the cylinder gasket and repeat the test when excessive loss is found.

Expected outcome: No unacceptable cylinder-side high-pressure leak is present.

7. Test the Flow-Control System

Action:

For units using the O₂ sensor, measured oxygen concentration must remain at least 25% throughout the O₂/N₂O linkage test.

For units tested without the O₂ sensor, verify the manufacturer-listed minimum O₂ flows:

Expected outcome: Gas controls, displays, floats, and the O₂/N₂O linkage operate normally.

8. Test the Vaporizer Mounting and Back Pressure

Use only vaporizers compatible with the installed manifold.

Action:

Pass criterion: O₂ flow does not decrease by more than 1 L/min over the tested vaporizer range.

Use approved scavenging or collection during this test.

Expected outcome: Vaporizer mounting, interlock, and machine back pressure are acceptable.

9. Test the O₂ Flush

Mechanical Ventilation Position

Action:

Pass criterion: Bellows fully inflate within 1–3 seconds.

Manual Ventilation Position

Action:

Pass criteria:

Expected outcome: O₂ flush flow is adequate and the pushbutton returns normally.

10. Inspect Breathing-System Components

Action: Check:

Confirm that the inspiratory valve opens during inspiration and closes at expiration. Confirm that the expiratory valve opens during expiration and closes at inspiration.

Expected outcome: The breathing system is correctly assembled, clean, undamaged, and mechanically functional.

11. Perform the Bellows Test

Action:

Pass criteria:

Expected outcome: The bellows assembly does not show an obvious leak.

12. Perform the Mechanical-Ventilation Leak Test

Action:

Pass criterion: The machine displays that the test passed, indicating leakage is within approximately 0.15–0.2 L/min.

Expected outcome: Bellows, drive-gas circuit, absorber, tubing, flow sensors, and connectors have acceptable airtightness.

13. Perform the Manual-Ventilation Leak Test

Action:

Pass criterion: No meaningful pressure decrease indicating a breathing-system leak.

Expected outcome: The manual circuit, APL valve, check valves, absorber, tubing, sensors, and connections are leak-free.

14. Test the APL Valve

Action:

Pass criteria:

Adjust APL-valve accuracy according to the service manual when outside limits.

Expected outcome: The APL valve regulates and relieves pressure correctly.

15. Test the ACGO Pressure-Relief Valve

Perform only when ACGO is configured and the required fixture and measurement equipment are available.

Pass criterion: Pressure at valve opening is 100–125 cmH₂O.

The service manual identifies pressure-relief valve assembly P/N 0621-30-69662 when replacement is necessary.

Expected outcome: The ACGO pressure-relief valve opens within the manufacturer-listed range.

16. Test Mechanical Ventilation

Connect a 2 L bag as a test lung.

Volume-Controlled Ventilation

Test the following manufacturer-listed combinations using fresh gas of approximately 0.5 L/min:

Pass criteria:

Pressure-Controlled Ventilation

Perform only when PCV is configured.

Manufacturer-listed test settings:

Pass criteria:

Expected outcome: The ventilator delivers and measures volume and pressure within the manufacturer-listed tolerances.

17. Test Alarms

Use the manual’s test setup unless facility policy provides a validated equivalent.

Verify operation and cancellation of:

Expected outcome: Each alarm activates under the simulated condition and clears when the condition is corrected.

18. Inspect the AGSS

Action:

The manual identifies an AGSS disposal-system range of approximately 50–80 L/min.

Expected outcome: Waste-gas transfer and receiving components are intact, unobstructed, and functional.

19. Test AC and Battery Operation

Action:

Expected outcome: The machine transfers between AC and battery operation without unintended shutdown.

Review the three-year battery replacement schedule and replace applicable batteries when due.

20. Check Sensor Zero Points

Access the manufacturer diagnostic display with the system in Standby, fresh gas off, and the patient connection open to atmosphere.

Manufacturer-listed ranges:

Expected outcome: Actual measured pressure and flow values read zero, and A/D values are within the listed ranges.

The manual states that an out-of-range zero point cannot be corrected through calibration and may indicate that the monitor board must be replaced.

21. Verify Flow-Sensor Accuracy and Calibrate

Using a calibrated flow reference, verify the sensors at approximately:

Pass criterion: Each machine flow-sensor reading agrees with the calibrated reference within 1 L/min or 5% of the reference reading, whichever is greater.

Perform the applicable manufacturer flow calibration when:

Expected outcome: Flow sensors meet manufacturer-listed accuracy and calibration completes successfully.

22. Verify Pressure-Sensor Accuracy and Calibrate

Using a calibrated pressure reference, verify at approximately:

Pass criterion: Airway and PEEP pressure readings agree with the reference within 1 cmH₂O or 2% of the reference reading, whichever is greater.

Perform pressure calibration after replacement of the monitor board, expiratory-valve assembly, or three-way valve assembly, or when the accuracy criterion is not met.

Expected outcome: Pressure sensors meet manufacturer-listed accuracy.

23. Verify Electronic Flowmeter Accuracy

Using an appropriate calibrated gas-flow reference:

Pass criterion: Electronic flowmeter indication agrees with the reference within 10% of the reference reading.

Calibrate when the electronic flowmeter board or throttling device has been replaced or when accuracy is outside the manufacturer-listed limit.

Expected outcome: Electronic gas-flow displays meet manufacturer-listed accuracy.

24. Calibrate the O₂ Sensor

Perform when the O₂ sensor is configured.

Calibrate at:

Calibration is manufacturer-listed after O₂ sensor or monitor-board replacement, or when measured oxygen concentration deviates by more than 3% in both room air and pure oxygen.

Expected outcome: O₂ calibration completes and the displayed value is appropriate in room air and oxygen.

25. Perform the Low-Pressure Pneumatic Leak Test

Perform the manufacturer low-pressure pipeline leak test referenced in the annual maintenance checklist.

Document:

Expected outcome: The low-pressure pneumatic system has no leak exceeding the manufacturer’s test criteria.

26. Complete Final Checkout

Action:

Expected outcome: The machine is fully assembled, configured, and ready for clinical pre-use checkout.

Electrical Safety Testing

The Mindray service manual specifies leakage current not greater than 500 µA and protective-earth impedance from the power-cord grounding terminal to exposed metal less than 0.2 Ω.

Ground Resistance

Ground resistance measured in ohms, manufacturer limit <0.2 Ω:

Result =

Pass / Fail =

Leakage Current

Touch/chassis leakage measured in microamps:

Facility limit used = 500 µA

Result =

Pass / Fail =

Document the electrical-safety analyzer, asset number, calibration due date, test standard or mode, applied-parts configuration, and power-cord condition according to facility policy.

Pass / Fail Criteria

PM Pass Criteria

Remove From Service Criteria

Remove the machine from service when any of the following remains unresolved:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide