Mortara ELI 250 / ELI 210

Preventive Maintenance Procedure

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

Mortara

Model

ELI 250 / ELI 210

Manual Reference

ELI 250 / ELI 210 Service Manual, REF 9516-163-50-ENG, Revision B1

Model scope: The provided service manual specifically covers the ELI 250 and ELI 210. Do not automatically apply every step in this procedure to other Mortara ELI Series models without confirming compatibility with the model-specific service documentation.

Manufacturer Recommendation / AEM Status

Manufacturer-Listed PM Frequency

Clean and inspect the unit every 6 months.

Increase cleaning and inspection to every 3 months when the unit is in heavy use.

Clean the printhead every 80 operating hours. The manual also recommends monthly printhead cleaning.

Perform leakage testing according to facility or regulatory requirements.

Periodically inspect the power cord and transmission cables for fraying or damage. The manual does not assign a specific interval to this cable inspection.

The customer is responsible for establishing and implementing a satisfactory maintenance schedule.

Post-Repair Safety Testing

Perform leakage testing after removal or replacement of the following assemblies:

The manual also requires leakage testing when these assemblies are removed or replaced during normal maintenance or repair.

Parts Relevant to Maintenance

The manual does not provide a scheduled battery-replacement interval.

AEM Status

If the facility performs this PM at a different interval or changes the manufacturer-listed activities, document the change under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Do not use alcohol, solvents, or cleaning solutions on the exterior covers.

Step-by-Step PM Procedure

1. Confirm Device Identification and PM Interval

Action: Record the model, serial number, asset number, software version when required by facility policy, and operating environment.

Confirm whether the device is maintained at a six-month interval, a three-month interval because of heavy use, or another interval approved under the facility’s AEM program.

Expected outcome: Equipment identification is legible and the applicable PM interval is documented.

2. Remove the Device From Clinical Use

Action: Confirm that no patient is connected. Disconnect accessories that are not required for testing and inspect the unit before applying power.

Do not disassemble the unit while AC power is connected. Remove the battery fuse before internal disassembly.

Expected outcome: The unit is safely prepared for inspection and testing.

3. Clean the Exterior

Action: Wipe the exterior covers and line cord with a cloth dampened with water. Dry all surfaces before applying power.

Do not submerge, autoclave, or steam-clean the device or patient cables. Do not use alcohol, solvents, or cleaning solutions on the exterior plastic surfaces. Do not use abrasive materials on labels or nameplates.

Expected outcome: Exterior surfaces are clean, dry, and free from residue or damage.

4. Perform Exterior Inspection

Action: Inspect the complete unit for:

Expected outcome: The enclosure and accessible components are secure and free from damage that could affect safety or operation.

5. Inspect the Power Cord and Communication Cables

Action: Inspect the hospital-grade power cord and installed transmission cables for cuts, fraying, crushed insulation, bent or loose contacts, discoloration, overheating, or damage at the connector and strain relief.

Confirm that the power cord is the correct three-wire grounded type.

Expected outcome: Cords and connectors are intact, properly grounded, and suitable for continued service.

Remove from service when: A conductor is exposed, the ground pin is missing, the strain relief is damaged, or a connector is loose or shows evidence of overheating.

6. Inspect the Patient Cable and Lead Wires

Action: Inspect the patient cable and lead wires for:

The manual states that compatible patient cables include at least 10 kΩ series resistance in each lead for defibrillation protection. Do not substitute an unapproved cable.

Expected outcome: The patient cable and lead wires are intact and produce stable connections.

7. Inspect and Clean the Printhead

Action: Open the writer cover and inspect the paper path, platen, cue-sensor area, and printhead.

Apply isopropyl alcohol to a clean cloth. Wipe the printhead until foreign material is removed, allow it to dry completely, and load the correct thermal paper.

Expected outcome: The printhead is clean and dry, and paper feeds without obstruction.

Manufacturer-listed interval: Every 80 operating hours, with monthly cleaning also recommended.

8. Verify AC and Battery Operation

Action: Connect the unit to an appropriate hospital-grade grounded outlet. Verify that the AC or charging indicator illuminates. Power the unit on and confirm normal startup.

Disconnect AC power and confirm that the device continues operating from its internal battery. Reconnect AC power after testing so the battery can recharge.

Expected outcome: The unit powers up from AC, transfers to battery operation, and indicates charging when AC power is restored.

The manual does not provide a specific minimum battery runtime for PM acceptance. Use facility policy or an approved battery-capacity procedure when runtime testing is required.

9. Test Keyboard Operation

Action: Enter a test-patient screen and exercise the alphanumeric keys, function keys, Enter, Backspace, Auto, Rhythm, Stop, and other controls used during normal operation.

Expected outcome: Each selected character or function is recognized without sticking, double-entry, or missed input.

10. Test the Display

Action: Apply an ECG simulator signal. Inspect the display for missing segments, gaps, black areas, flickering, distortion, or intermittent operation.

Adjust contrast through its normal range when accessible, then return it to a usable setting.

Expected outcome: The display is readable and presents complete waveforms and menu information.

11. Test the Speaker

Action: Access the configured volume setting and verify that the speaker produces an audible tone. Change the volume through multiple settings and confirm that the tone changes appropriately from setting 0 through setting 8.

Expected outcome: The speaker operates and responds correctly to volume changes.

12. Verify Date and Time

Action: Open the Date/Time screen and confirm that the displayed date and time are correct. Correct them according to facility policy when necessary.

Restart the unit and verify that the date and time are retained.

Expected outcome: The device displays and retains the correct date and time.

13. Perform the Writer Test

Action: With AC power connected, initiate the writer test using the applicable service key sequence:

ALT + SHIFT + RHY

Inspect the test printout for:

Measure the 200 mm reference line with a metric ruler.

Expected outcome:

The manual notes that the printed circle may show a gap when the configured filter is below 300 Hz. This condition alone should not be treated as a writer failure.

14. Verify Calibration Pulse and Paper Speed

Action: Print an ECG at the paper speeds used by the facility. Measure the calibration pulse with a metric ruler rather than relying on the printed paper grid.

Expected calibration-pulse width:

Expected calibration-pulse amplitude:

Expected outcome: Measured pulse width and amplitude correspond to the selected speed and gain.

15. Perform ECG Acquisition Test

Action: Connect the ECG simulator to the patient cable and apply a known 12-lead ECG signal.

Enter test-patient information, acquire an automatic ECG, print the ECG when printing is not automatic, and compare the displayed and printed ECG with the simulator settings.

Expected outcome:

The manual does not state one universal amplitude, rate, or interval tolerance for this test. Use the simulator setting and facility-approved verification criteria.

16. Perform Lead-Fail Test

Action: With the patient cable attached to a simulator or approved shorting device, disconnect one lead at a time.

Verify the appropriate message:

Expected outcome: Only the appropriate lead-fail indication appears for each disconnected lead.

17. Optional Noise Test

Action: Perform this test only when the required shorting plug and an approved facility procedure are available.

Connect the specified shorting plug, set gain to 20 mm/mV, print a rhythm strip, allow at least 200 mm of paper to print, and inspect the baseline.

Expected outcome: No baseline peak exceeds 0.5 mm.

The manual states that Mortara does not supply the shorting plug.

18. Inspect Installed Communication Functions

Action: When communication features are part of the device’s clinical use, verify the applicable installed interface:

Confirm that configuration values have not been unintentionally changed. Perform a test transmission when included in the facility PM procedure.

Expected outcome: The installed communication method connects and transmits successfully, or the limitation is documented and referred to the appropriate IT or system-support group.

19. Optional Internal Inspection

Perform internal inspection only when included in facility policy and completed by qualified service personnel.

Action: Remove AC power and the battery fuse before opening the unit.

Inspect for:

Expected outcome: Internal components are secure, clean, and free from damage.

After reinstalling the PCB assembly, AC connector assembly, or writer assembly, perform the manufacturer-listed leakage testing.

Electrical Safety Testing

The ELI 250 and ELI 210 are identified as Class I equipment with Type CF applied parts.

The service manual references ANSI/AAMI ES-1 and IEC 60601-1 and identifies the following safety tests:

The manual does not provide measured leakage-current limits. Use the current facility-approved standard and test configuration.

Protective-Earth Resistance

This acceptance value is based on facility policy rather than a numerical limit stated in the provided manual.

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:

Result =

Touch/Chassis Leakage

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Patient Leakage

Patient-source leakage measured in microamps =

Facility limit used =

Pass / Fail =

Patient-sink leakage measured in microamps =

Facility limit used =

Pass / Fail =

Post-Repair Requirement

Complete leakage testing after removal or replacement of the PCB assembly, AC connector assembly, or writer assembly.

Pass / Fail Criteria

Pass

The device may pass PM when:

Remove From Service

Remove the unit from service for:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide