Philips Avalon FM30

Preventive Maintenance Procedure

On this page

Asset Type

Fetal Monitor

Manufacturer

Philips

Model

Avalon FM30

Manual Reference

Avalon Fetal Monitor FM20/FM30 Service Guide, Part Number M2703-9000A, printed 07/05.

Manufacturer Recommendation / AEM Status

Philips separates the manufacturer-listed activities into preventive maintenance, other regular tests, performance assurance tests, and safety tests.

Manufacturer-Listed Preventive Maintenance

The manual states that preventive maintenance is restricted to noninvasive blood pressure calibration and associated performance testing.

NIBP calibration and performance testing: Once every two years, or as specified by local laws, whichever comes first.

The FM30 must be equipped with the optional NIBP assembly for these tests to apply.

Manufacturer-Listed Other Regular Tests

Visual inspection: Before each use.

Recorder maintenance: Once a year, or when print quality becomes degraded.

Transducer and patient-module testing: Once a year, or when a measurement is suspected to be incorrect.

Manufacturer-Listed Performance Assurance

NIBP performance tests: Once every two years, or when the measurement is suspected to be incorrect.

SpO₂ performance test: Once every two years, or when the measurement is suspected to be incorrect.

Manufacturer-Listed Electrical Safety Testing

Perform enclosure and patient leakage-current tests:

The FM30 is a Class II device. Protection against electric shock is provided by double or reinforced insulation. The protective-earth conductor is used for electromagnetic compatibility and is not the primary means of protection against electric shock.

A conventional protective-earth or ground-resistance test is therefore not the manufacturer-listed electrical safety test for this monitor.

Manufacturer-Listed Post-Repair Testing

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

IUP Test Equipment When Included

Step-by-Step PM Procedure

1. Identify the Monitor and Installed Options

Action: Record the model, serial number, software revision, and installed options. Determine whether the monitor includes NIBP, SpO₂, LAN/RS232 interface, ultrasound transducers, Toco or Toco+ transducers, and an ECG/IUP patient module.

Expected outcome: The device and all applicable PM test sections are correctly identified.

2. Review Service History

Action: Review prior PM results, repairs, impact-damage reports, recurring recorder problems, transducer failures, and previous electrical safety values.

Expected outcome: Any testing required because of repair history or impact damage is added to the work scope.

3. Perform the Visual Inspection

Action: Inspect the following:

Replace seriously damaged cables immediately. Do not use a transducer with a broken or damaged belt button.

Expected outcome: No cracks, loose parts, damaged cables, damaged sockets, or insecure mounting components are present.

4. Perform the Power-On Self-Test

Action: Connect the monitor to AC power and turn it on.

Verify that:

Expected outcome: The monitor completes its startup self-test and reaches the monitoring screen without an unresolved fault.

5. Check the Display, Touchscreen, Controls, and Alarm Audio

Action: Confirm the display is readable and has normal brightness. Verify touchscreen response across the active screen area. Confirm that keys and controls respond correctly and that audible alarm and system sounds are available.

Confirm the speaker is not muted or set to zero solely for testing. Calibrate the touchscreen only when the touch position is inaccurate.

Expected outcome: Display, touch input, controls, and audible output operate normally.

6. Inspect and Clean the Fetal Recorder

Action: Open the paper drawer and inspect the recorder.

Expected outcome: The platen, printhead, and sensor are clean, and the paper is correctly installed.

7. Run the Fetal Recorder Self-Test

Action: In Service Mode, select:

Main Setup → Fetal Recorder → Selftest

Inspect the printed test pattern.

Manufacturer-listed criteria:

Expected outcome: The recorder feeds paper consistently and meets the printhead and alignment criteria.

If the resting Toco trace does not print at 20 units, perform recorder calibration and adjust the offset.

8. Test the Ultrasound Transducers

Action: For each ultrasound transducer:

Expected outcome: A sound is produced from each of the seven crystal positions and when the transducer is moved over a solid surface.

If the test fails, repeat it using another transducer to distinguish a transducer failure from a monitor or connector-block problem.

9. Test the Toco Transducers

Action:

Expected outcome:

10. Test Toco+ or Patient-Module Functions When Included

DECG Mode

Action: Connect the DECG adapter cable M1362B. Confirm that the monitor initially displays the DECG LEADS OFF INOP. Connect a fetal scalp electrode and establish continuity between the spiral and reference electrode using an approved test method.

Expected outcome: The DECG LEADS OFF INOP disappears.

MECG Mode

Action: Connect the MECG adapter cable M1363A and attach it to electrodes or a patient simulator.

Expected outcome: MECG values appear on the maternal display or recorder trace.

IUP Mode

Action: Apply 80 mmHg ±5 mmHg using the specified manometer, expansion chamber, tubing, T-adapter, IUP cable, and catheter setup. Slowly release the applied pressure.

Expected outcome: The monitor and printed trace correspond to the applied pressure and decrease as pressure is released.

11. Perform the NIBP Accuracy Test

Applicable only when NIBP is installed.

Action: Connect the reference manometer, hand pump, tubing, and 250 mL expansion chamber.

In Service Mode:

Expected outcome: The difference between the monitor and reference manometer is ≤3 mmHg.

If the difference exceeds 3 mmHg, perform the manufacturer-listed NIBP calibration procedure.

12. Perform the NIBP Leakage Test

Action: With the system at approximately 280 mmHg, record the initial pressure, wait 60 seconds, record the final pressure, and calculate the pressure loss.

Expected outcome: Pressure loss is less than 6 mmHg in 60 seconds.

Result calculation: Initial pressure − Final pressure

13. Perform the NIBP Linearity Test

Action: Reduce the pressure to 150 mmHg. Wait 10 seconds and compare the monitor value with the reference manometer.

Expected outcome: The difference is ≤3 mmHg.

14. Perform the NIBP Valve Test

Action: Raise the pressure to 280 mmHg. Set Close Valves to Off. Wait five seconds and record the remaining displayed pressure.

Expected outcome: The remaining pressure is less than 10 mmHg.

15. Perform the SpO₂ Performance Test

Applicable only when SpO₂ is installed.

Action: Connect an adult SpO₂ sensor and obtain a reading on a healthy test subject according to facility policy.

Expected outcome: Displayed SpO₂ is 95%–100%.

This is the manufacturer-listed functional check in the provided service guide. It is not a substitute for a simulator-based accuracy test when one is required by facility policy.

Electrical Safety Testing

The manufacturer-listed periodic safety tests for the Avalon FM30 are leakage-current tests.

16. Enclosure and Functional-Earth Leakage Test

Action: Measure the highest leakage current in normal and reversed polarity and under normal and single-fault conditions using the manufacturer-described accessible metal points.

Manufacturer-listed limit: Sum of functional-earth and enclosure leakage current ≤100 µA.

Result =

17. Patient Leakage-Current Test

Applicable to ECG measurement inputs.

Action: Perform the patient leakage-current test with mains applied to the applied part under a single-fault condition at the specified test voltage.

Manufacturer-listed limit: Patient leakage current ≤50 µA at 264 V.

Result =

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Ground resistance is not included as a manufacturer-listed PM measurement because the FM30 is a Class II, double-insulated device. Document any additional facility-required electrical safety testing separately.

18. Complete the Final Functional Verification

Action: Confirm the following:

Expected outcome: The monitor is ready for clinical use with no unresolved safety, measurement, recorder, or accessory defect.

Pass / Fail Criteria

Pass

Remove From Service

Remove the monitor or affected accessory from service when:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide