Philips IntelliVue MX Series

Preventive Maintenance Procedure

On this page

Asset Type

Patient Monitor

Manufacturer

Philips

Model

IntelliVue MX Series

Manual Reference

IntelliVue Patient Monitor MX400/MX450/MX500/MX550 Service Guide, Release K.1, Part Number 453564436301

Manufacturer Recommendation / AEM Status

The service guide separates preventive maintenance, recurring safety testing, performance assurance testing, and pre-use checks.

Manufacturer-Listed Frequencies

The manufacturer specifically identifies NBP calibration and Microstream CO₂ calibration as preventive-maintenance procedures. The other listed inspections, safety tests, and performance tests are recurring maintenance activities that support the overall PM program.

This procedure focuses primarily on the MX400, MX450, MX500, and MX550 display monitor. Measurement testing should also be completed for the MMS, IntelliVue X2, extensions, parameter modules, interfaces, and optional accessories installed with the monitor.

No manufacturer-listed calendar replacement interval was found for the monitor battery. Battery replacement should be based on condition, performance, diagnostic indications, service history, and facility policy.

After maintenance or repair, the service guide states that the monitor must not be returned to operation until the applicable safety and performance tests have been completed. Before handoff, confirm that the monitor is appropriately configured, placed in Monitoring Mode, and supplied with the current user documentation.

AEM Status

If a facility performs this PM at a different interval or changes the listed activities, the change should be managed under the facility’s approved AEM program, documented risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Not every monitor configuration requires every listed tool.

Step-by-Step PM Procedure

1. Identify the Installed Configuration

Record the following:

Expected outcome: The PM scope matches the actual monitor configuration.

2. Remove the Monitor From Clinical Use

Remove the monitor from active patient use before testing.

Disconnect patient leads, sensors, transducers, and sampling lines. Decontaminate the monitor and accessories according to facility policy and the Philips Instructions for Use.

Expected outcome: The monitor can be tested without placing a patient or operator at risk.

3. Perform the External Visual Inspection

Inspect the following:

After service or repair, verify that no loose hardware, debris, or foreign material remains inside or around the monitor.

Expected outcome: No damage, contamination, insecure mounting, loose parts, or compromised connectors are present.

Do not return to service: Cracked housing, damaged display, liquid ingress, damaged mains components, exposed conductors, swollen battery, bent connector contacts, insecure mounting, or any condition that could affect safety or reliable operation.

4. Inspect the Mounting System

Confirm that:

The manufacturer-listed mounting integrity test is particularly applicable when the quick mount has been removed, reassembled, found loose, or is unstable.

Expected outcome: The mount securely supports the monitor without excessive movement or damaged hardware.

5. Perform the Power-On Test

  1. Connect the monitor to verified AC power.
  2. Install and connect the normal MMS, IntelliVue X2, modules, extensions, displays, and interfaces included in the PM.
  3. Power on the monitoring system.
  4. Confirm successful completion of the startup sequence.
  5. Verify that no unresolved startup error or technical INOP remains.
  6. Confirm that the monitoring screen loads.
  7. Confirm that an ECG waveform area appears. A flat line is acceptable when no simulator is connected.
  8. Verify that installed modules and configured interfaces are recognized.

Expected outcome: The monitor completes startup and enters normal operation without unresolved errors.

6. Check the Display, Touchscreen, and Operator Controls

Verify that:

Calibrate the touchscreen only when touch accuracy is incorrect or when required following display-related service.

Expected outcome: The user interface operates consistently and accurately.

7. Verify Audible and Visual Alarm Operation

Generate or simulate an alarm condition.

Confirm that:

Expected outcome: Local audible and visual alarm indications operate reliably.

8. Test AC Power and Battery Transfer

  1. Confirm that the monitor indicates connection to AC power.
  2. Confirm that the battery is recognized and charging.
  3. With the monitor operating, disconnect AC power.
  4. Verify continued monitor operation on battery without an unintended restart.
  5. Confirm that the power-loss alarm buzzer sounds. The manufacturer-listed expected result is approximately one minute unless AC power is restored or the power button is pressed.
  6. Reconnect AC power.
  7. Confirm return to AC operation and charging.

When facility policy includes capacity testing, document the battery runtime or analyzer result.

Expected outcome: The monitor transfers between AC and battery power correctly, the battery remains recognized, and the power-loss alarm operates.

Do not return to service: Unexpected shutdown, battery not recognized, battery swelling or leakage, failure to charge, unreliable transfer, or inadequate runtime under facility criteria.

9. Perform Basic Performance Assurance

Place the monitor in Demo Mode or connect the appropriate simulator.

Confirm that each installed measurement device displays values, including applicable:

Expected outcome: Every installed and connected parameter is recognized and displays data.

A full parameter performance test is manufacturer-listed once every two years or whenever a measurement is suspected to be incorrect.

10. Perform NBP Performance Testing

Complete this section when an NBP-capable MMS or IntelliVue X2 is included in the PM.

Accuracy at 280 mmHg

  1. Connect the reference manometer, hand pump, valve, tubing, and 250 mL ±10% expansion chamber.
  2. Enter Service Mode and open the NBP service menu.
  3. Set Close Valves to On.
  4. Apply 280 mmHg.
  5. Allow ten seconds for stabilization.
  6. Compare the displayed pressure with the reference manometer.

Expected result:

Calibrate NBP when the difference exceeds 3 mmHg.

Leakage

  1. Maintain approximately 280 mmHg.
  2. Observe the pressure for 60 seconds.
  3. Calculate the pressure loss.

Expected result: Pressure loss less than 6 mmHg in 60 seconds.

Linearity

  1. Reduce pressure to 150 mmHg.
  2. Allow ten seconds for stabilization.
  3. Compare the monitor reading with the reference manometer.

Expected result:

Valve Operation

  1. Raise pressure to 280 mmHg.
  2. Set Close Valves to Off.
  3. Wait five seconds.
  4. Record the remaining pressure.

Expected result: Remaining pressure less than 10 mmHg.

11. Perform Configured Parameter Checks

Use only the checks applicable to the installed measurement hardware.

ECG

Set the patient simulator to sinus rhythm at 100 or 120 bpm.

Expected result: 100 bpm ±2 bpm or 120 bpm ±2 bpm.

Respiration

Set the simulator to:

Expected result: 40 rpm ±2 rpm or 45 rpm ±2 rpm.

SpO₂

The service guide describes a functional check using an adult sensor on a healthy person.

Expected result: 95% to 100%.

Facility policy may instead require an approved pulse-oximeter simulator. A functional simulator verifies response to its test curve; it does not independently establish clinical SpO₂ accuracy.

Invasive Pressure

  1. Apply zero pressure and zero the channel.
  2. Set the simulator to 200 mmHg.

Expected result: 200 mmHg ±5 mmHg.

Temperature

Set the simulator to 40°C or 100°F.

Expected result: 40°C ±0.2°C or 100°F ±0.4°F.

ECG Output

When ECG synchronization output is used, connect a simulator and compatible defibrillator or receiving device.

Expected result: The ECG signal is displayed by the receiving device.

CO₂

Complete the manufacturer-listed accuracy, flow, calibration, and operating-hour-based maintenance procedures applicable to the installed mainstream, sidestream, or Microstream CO₂ hardware.

Expected outcome: Applicable CO₂ checks pass the limits and procedures specified for the installed measurement hardware.

12. Test Configured Interfaces and Accessories

Complete only the checks applicable to interfaces and accessories that are installed or used.

Integrated Recorder

Start a recording.

Expected outcome: The recording prints correctly and is readable.

Nurse Call

Verify that the relay changes state when an alarm is generated. Use the correct pinout for the installed standard or flexible nurse-call interface.

Expected outcome: Correct configured relay activity occurs during the alarm condition.

IntelliBridge

Connect the supported external device and verify that required numerics or waveforms appear on the monitor.

Expected outcome: External-device information is displayed correctly.

Wired Network

Confirm that the LAN connection is recognized and that the monitor communicates with the intended Philips network or central station according to facility policy.

Wireless Network

Confirm connection status, signal strength, and communication according to the facility’s validated wireless configuration. Do not change production wireless settings merely to perform routine PM.

Remote Control or SRR

Confirm that the assigned remote control operates the monitor.

Electrical Safety Testing

Test the monitor as a standalone mains-powered device. Include detachable power cords and accessories that can affect the test result.

Disconnect data cables or other paths that may create an unintended earth connection unless they are specifically included as part of a medical electrical system test.

Protective Earth Resistance

Measure from the protective-earth terminal to the equipotential connector or another protectively earthed accessible metal point. Flex the power cord while testing.

Manufacturer-listed limit with mains cable: ≤0.300 Ω

Ground resistance measured in ohms, expected ≤0.300 Ω or facility/NFPA 99 criteria if different:

Result =

Testing must stop immediately if the protective-earth test fails.

Equipment Leakage Current — Normal Condition

Test normal and reversed polarity using the direct method described in the service guide.

Touch/chassis leakage measured in microamps:

Facility limit used = ≤100 µA

Result =

Pass / Fail =

Equipment Leakage Current — Single-Fault Condition

Open protective earth as directed by the electrical safety analyzer procedure.

Touch/chassis leakage measured in microamps:

Facility limit used = ≤300 µA

Result =

Pass / Fail =

Applied-Part Leakage

Test each applicable applied-part group separately. Multiple connections within an applied part, such as ECG leads, are tested together.

The service guide warns against simultaneously connecting invasive-pressure and temperature interfaces during the applied-part test because they share an internal insulation interface.

Manufacturer-listed CF limit, mains on applied part: ≤50 µA

Applied-part leakage measured in microamps:

Applied part tested =

Manufacturer limit used =

Result =

Pass / Fail =

Medical Electrical System Test

Complete a system test when the monitor forms a medical electrical system with connected displays, IT equipment, external devices, or a multiple socket outlet.

Manufacturer-listed limits include:

Document all safety results for trending.

13. Restore and Finalize the Monitor

  1. Reinstall all normal accessories and modules.
  2. Confirm that the monitor is in Monitoring Mode, not Service Mode or Demo Mode.
  3. Confirm that the correct profile and facility configuration are active.
  4. Clear test patient information according to facility policy.
  5. Confirm the correct date and time.
  6. Verify alarm settings or default profile status according to facility procedure.
  7. Confirm that no unresolved technical alarms, error messages, or service prompts remain.
  8. Clean the device and prepare it for clinical use.

Expected outcome: The complete configured monitoring system is safe, functional, correctly configured, and ready for use.

Pass / Fail Criteria

Pass

The monitor passes when:

Remove From Service

Remove the monitor from service when:

The service guide directs testing to be discontinued immediately following a failed test until the device is repaired or identified as defective.

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide