On this page
Asset Type
Manufacturer
Model
Manual Reference
IntelliVue Patient Monitor MX400/MX450/MX500/MX550 Service Guide, Release K.1, Part Number 453564436301
Manufacturer Recommendation / AEM Status
The service guide separates preventive maintenance, recurring safety testing, performance assurance testing, and pre-use checks.
Manufacturer-Listed Frequencies
- Visual inspection: Before each use and after each service, maintenance, or repair event.
- Power-on test: Before each use.
- Protective earth resistance: Once every two years and after repairs involving the power supply, mains socket, protective-earth connection, or impact damage.
- Equipment leakage current: Once every two years and following the applicable repair conditions.
- Applied-part leakage current: Once every two years and following the applicable repair conditions.
- System safety test: Once every two years when applicable to the installed medical electrical system.
- Performance assurance tests: Once every two years, or whenever a measurement is suspected to be incorrect.
- Mainstream and sidestream CO₂ accuracy and flow checks: Once every year.
- NBP performance and calibration: At least once every two years, or more often when required by local requirements.
- Microstream CO₂ calibration: Once every year or after 4,000 hours of continuous operation, and after repair or replacement of instrument parts.
- M3015A old hardware Revision A pump and scrubber: Every three years or after 15,000 operating hours.
The manufacturer specifically identifies NBP calibration and Microstream CO₂ calibration as preventive-maintenance procedures. The other listed inspections, safety tests, and performance tests are recurring maintenance activities that support the overall PM program.
This procedure focuses primarily on the MX400, MX450, MX500, and MX550 display monitor. Measurement testing should also be completed for the MMS, IntelliVue X2, extensions, parameter modules, interfaces, and optional accessories installed with the monitor.
No manufacturer-listed calendar replacement interval was found for the monitor battery. Battery replacement should be based on condition, performance, diagnostic indications, service history, and facility policy.
After maintenance or repair, the service guide states that the monitor must not be returned to operation until the applicable safety and performance tests have been completed. Before handoff, confirm that the monitor is appropriately configured, placed in Monitoring Mode, and supplied with the current user documentation.
AEM Status
If a facility performs this PM at a different interval or changes the listed activities, the change should be managed under the facility’s approved AEM program, documented risk assessment, maintenance history, and applicable regulatory requirements.
What This PM Procedure Covers
- Monitor housing, display, touchscreen, controls, connectors, and mount
- Startup, self-test, display, alarm indicators, and audible alarm operation
- AC power, battery transfer, charging indication, and power-loss alarm
- Electrical safety testing
- Basic performance assurance
- Configured interface and accessory checks
- Applicable MMS, IntelliVue X2, parameter-module, NBP, and CO₂ testing
Required Tools and Equipment
- Electrical safety analyzer compliant with facility requirements
- Patient simulator for configured physiological parameters
- Reference NBP manometer accurate to 0.2% of reading
- Hand pump, valve, appropriate tubing, and 250 mL ±10% expansion chamber
- Temperature simulator with 0.1°C or 0.2°F resolution when temperature is configured
- Ohmmeter when nurse-call testing is applicable
- Philips-approved power cord and configured accessories
- Cleaning materials specified by facility policy and the Philips Instructions for Use
- CO₂ test equipment and calibration gases when CO₂ testing is applicable
- Facility-approved load or battery test equipment when battery capacity testing is included by policy
Not every monitor configuration requires every listed tool.
Step-by-Step PM Procedure
1. Identify the Installed Configuration
Record the following:
- Exact model: MX400, MX450, MX500, or MX550
- Serial number
- Software revision
- Installed battery
- MMS or IntelliVue X2
- Parameter modules and extensions
- Integrated recorder
- Wired or wireless networking
- Nurse-call interface
- IntelliBridge interface
- Remote control or SRR
- External display
- Installed mounting system
Expected outcome: The PM scope matches the actual monitor configuration.
2. Remove the Monitor From Clinical Use
Remove the monitor from active patient use before testing.
Disconnect patient leads, sensors, transducers, and sampling lines. Decontaminate the monitor and accessories according to facility policy and the Philips Instructions for Use.
Expected outcome: The monitor can be tested without placing a patient or operator at risk.
3. Perform the External Visual Inspection
Inspect the following:
- Front and rear housings for cracks, impact damage, separation, or missing parts
- Display for cracks, discoloration, unreadable areas, or contamination
- Touchscreen surface
- Power switch and external controls
- Alarm lamps and ambient-light sensor area
- AC inlet, equipotential connector, LAN, USB, video, MSL, nurse-call, and other installed connectors
- MSL connector pins for bending, displacement, or breakage
- Power cord and strain relief for cuts, cracks, twisting, exposed conductors, or damaged pins
- Battery compartment and battery for damage, swelling, leakage, or contamination
- Recorder door and paper path when installed
- Installed module rack and module retention
- Labels and serial-number identification
- Mount, quick-release assembly, hardware, and locking mechanism
After service or repair, verify that no loose hardware, debris, or foreign material remains inside or around the monitor.
Expected outcome: No damage, contamination, insecure mounting, loose parts, or compromised connectors are present.
Do not return to service: Cracked housing, damaged display, liquid ingress, damaged mains components, exposed conductors, swollen battery, bent connector contacts, insecure mounting, or any condition that could affect safety or reliable operation.
4. Inspect the Mounting System
Confirm that:
- The monitor is securely attached to the approved mount.
- The release mechanism operates correctly.
- Mounting screws are tight.
- There is no abnormal clearance between the quick mount and monitor housing.
- The mounting threads and chassis attachment points are intact.
- The monitor does not shift, tilt, or release unintentionally.
The manufacturer-listed mounting integrity test is particularly applicable when the quick mount has been removed, reassembled, found loose, or is unstable.
Expected outcome: The mount securely supports the monitor without excessive movement or damaged hardware.
5. Perform the Power-On Test
- Connect the monitor to verified AC power.
- Install and connect the normal MMS, IntelliVue X2, modules, extensions, displays, and interfaces included in the PM.
- Power on the monitoring system.
- Confirm successful completion of the startup sequence.
- Verify that no unresolved startup error or technical INOP remains.
- Confirm that the monitoring screen loads.
- Confirm that an ECG waveform area appears. A flat line is acceptable when no simulator is connected.
- Verify that installed modules and configured interfaces are recognized.
Expected outcome: The monitor completes startup and enters normal operation without unresolved errors.
6. Check the Display, Touchscreen, and Operator Controls
Verify that:
- The display is bright, stable, and readable.
- No persistent flicker, lines, major dead areas, or image distortion are present.
- The touchscreen responds across the usable display area.
- Menus open and selections register correctly.
- Installed hard keys, remote control, mouse, keyboard, or barcode scanner respond when included in the configuration.
- Automatic or manual brightness control operates appropriately.
- Date and time are correct or synchronize according to facility configuration.
Calibrate the touchscreen only when touch accuracy is incorrect or when required following display-related service.
Expected outcome: The user interface operates consistently and accurately.
7. Verify Audible and Visual Alarm Operation
Generate or simulate an alarm condition.
Confirm that:
- The audible alarm is clearly heard and is not weak, distorted, or intermittent.
- Alarm lamps illuminate with the appropriate visual indication.
- An alarm message appears on the display.
- Alarm silence, pause, and acknowledgment controls respond correctly.
- Alarm volume cannot be unintentionally left at an unsafe level under the approved configuration.
- Nurse-call or central alarm transmission operates when included in the facility’s PM scope.
Expected outcome: Local audible and visual alarm indications operate reliably.
8. Test AC Power and Battery Transfer
- Confirm that the monitor indicates connection to AC power.
- Confirm that the battery is recognized and charging.
- With the monitor operating, disconnect AC power.
- Verify continued monitor operation on battery without an unintended restart.
- Confirm that the power-loss alarm buzzer sounds. The manufacturer-listed expected result is approximately one minute unless AC power is restored or the power button is pressed.
- Reconnect AC power.
- Confirm return to AC operation and charging.
When facility policy includes capacity testing, document the battery runtime or analyzer result.
Expected outcome: The monitor transfers between AC and battery power correctly, the battery remains recognized, and the power-loss alarm operates.
Do not return to service: Unexpected shutdown, battery not recognized, battery swelling or leakage, failure to charge, unreliable transfer, or inadequate runtime under facility criteria.
9. Perform Basic Performance Assurance
Place the monitor in Demo Mode or connect the appropriate simulator.
Confirm that each installed measurement device displays values, including applicable:
- MMS or IntelliVue X2
- ECG and respiration
- SpO₂
- NBP
- Invasive pressure
- Temperature
- CO₂
- Cardiac output
- Other installed parameter modules
- IntelliBridge-connected devices
Expected outcome: Every installed and connected parameter is recognized and displays data.
A full parameter performance test is manufacturer-listed once every two years or whenever a measurement is suspected to be incorrect.
10. Perform NBP Performance Testing
Complete this section when an NBP-capable MMS or IntelliVue X2 is included in the PM.
Accuracy at 280 mmHg
- Connect the reference manometer, hand pump, valve, tubing, and 250 mL ±10% expansion chamber.
- Enter Service Mode and open the NBP service menu.
- Set Close Valves to On.
- Apply 280 mmHg.
- Allow ten seconds for stabilization.
- Compare the displayed pressure with the reference manometer.
Expected result:
- Monitor reading: 280 mmHg ±3 mmHg
- Difference from reference: ≤3 mmHg
Calibrate NBP when the difference exceeds 3 mmHg.
Leakage
- Maintain approximately 280 mmHg.
- Observe the pressure for 60 seconds.
- Calculate the pressure loss.
Expected result: Pressure loss less than 6 mmHg in 60 seconds.
Linearity
- Reduce pressure to 150 mmHg.
- Allow ten seconds for stabilization.
- Compare the monitor reading with the reference manometer.
Expected result:
- Monitor reading: 150 mmHg ±3 mmHg
- Difference from reference: ≤3 mmHg
Valve Operation
- Raise pressure to 280 mmHg.
- Set Close Valves to Off.
- Wait five seconds.
- Record the remaining pressure.
Expected result: Remaining pressure less than 10 mmHg.
11. Perform Configured Parameter Checks
Use only the checks applicable to the installed measurement hardware.
ECG
Set the patient simulator to sinus rhythm at 100 or 120 bpm.
Expected result: 100 bpm ±2 bpm or 120 bpm ±2 bpm.
Respiration
Set the simulator to:
- Base impedance: 1,500 Ω
- Delta impedance: 0.5 Ω
- Respiration rate: 40 or 45 rpm
Expected result: 40 rpm ±2 rpm or 45 rpm ±2 rpm.
SpO₂
The service guide describes a functional check using an adult sensor on a healthy person.
Expected result: 95% to 100%.
Facility policy may instead require an approved pulse-oximeter simulator. A functional simulator verifies response to its test curve; it does not independently establish clinical SpO₂ accuracy.
Invasive Pressure
- Apply zero pressure and zero the channel.
- Set the simulator to 200 mmHg.
Expected result: 200 mmHg ±5 mmHg.
Temperature
Set the simulator to 40°C or 100°F.
Expected result: 40°C ±0.2°C or 100°F ±0.4°F.
ECG Output
When ECG synchronization output is used, connect a simulator and compatible defibrillator or receiving device.
Expected result: The ECG signal is displayed by the receiving device.
CO₂
Complete the manufacturer-listed accuracy, flow, calibration, and operating-hour-based maintenance procedures applicable to the installed mainstream, sidestream, or Microstream CO₂ hardware.
Expected outcome: Applicable CO₂ checks pass the limits and procedures specified for the installed measurement hardware.
12. Test Configured Interfaces and Accessories
Complete only the checks applicable to interfaces and accessories that are installed or used.
Integrated Recorder
Start a recording.
Expected outcome: The recording prints correctly and is readable.
Nurse Call
Verify that the relay changes state when an alarm is generated. Use the correct pinout for the installed standard or flexible nurse-call interface.
Expected outcome: Correct configured relay activity occurs during the alarm condition.
IntelliBridge
Connect the supported external device and verify that required numerics or waveforms appear on the monitor.
Expected outcome: External-device information is displayed correctly.
Wired Network
Confirm that the LAN connection is recognized and that the monitor communicates with the intended Philips network or central station according to facility policy.
Wireless Network
Confirm connection status, signal strength, and communication according to the facility’s validated wireless configuration. Do not change production wireless settings merely to perform routine PM.
Remote Control or SRR
Confirm that the assigned remote control operates the monitor.
Electrical Safety Testing
Test the monitor as a standalone mains-powered device. Include detachable power cords and accessories that can affect the test result.
Disconnect data cables or other paths that may create an unintended earth connection unless they are specifically included as part of a medical electrical system test.
Protective Earth Resistance
Measure from the protective-earth terminal to the equipotential connector or another protectively earthed accessible metal point. Flex the power cord while testing.
Manufacturer-listed limit with mains cable: ≤0.300 Ω
Ground resistance measured in ohms, expected ≤0.300 Ω or facility/NFPA 99 criteria if different:
Result =
Testing must stop immediately if the protective-earth test fails.
Equipment Leakage Current — Normal Condition
Test normal and reversed polarity using the direct method described in the service guide.
Touch/chassis leakage measured in microamps:
Facility limit used = ≤100 µA
Result =
Pass / Fail =
Equipment Leakage Current — Single-Fault Condition
Open protective earth as directed by the electrical safety analyzer procedure.
Touch/chassis leakage measured in microamps:
Facility limit used = ≤300 µA
Result =
Pass / Fail =
Applied-Part Leakage
Test each applicable applied-part group separately. Multiple connections within an applied part, such as ECG leads, are tested together.
The service guide warns against simultaneously connecting invasive-pressure and temperature interfaces during the applied-part test because they share an internal insulation interface.
Manufacturer-listed CF limit, mains on applied part: ≤50 µA
Applied-part leakage measured in microamps:
Applied part tested =
Manufacturer limit used =
Result =
Pass / Fail =
Medical Electrical System Test
Complete a system test when the monitor forms a medical electrical system with connected displays, IT equipment, external devices, or a multiple socket outlet.
Manufacturer-listed limits include:
- System equipment leakage, normal condition: ≤100 µA
- System equipment leakage, single-fault condition: ≤300 µA
- Protective-earth leakage of a multiple socket outlet: ≤300 µA
- Standalone and connected-system readings should not differ by more than ±10%.
Document all safety results for trending.
13. Restore and Finalize the Monitor
- Reinstall all normal accessories and modules.
- Confirm that the monitor is in Monitoring Mode, not Service Mode or Demo Mode.
- Confirm that the correct profile and facility configuration are active.
- Clear test patient information according to facility policy.
- Confirm the correct date and time.
- Verify alarm settings or default profile status according to facility procedure.
- Confirm that no unresolved technical alarms, error messages, or service prompts remain.
- Clean the device and prepare it for clinical use.
Expected outcome: The complete configured monitoring system is safe, functional, correctly configured, and ready for use.
Pass / Fail Criteria
Pass
The monitor passes when:
- Housing, display, connectors, cord, battery, and mount are intact.
- Startup completes normally.
- Display, touchscreen, controls, alarms, and indicators operate correctly.
- AC and battery transfer operate without interruption.
- Installed modules and interfaces are recognized.
- Applicable parameter results remain within manufacturer-listed tolerances.
- Protective earth is ≤0.300 Ω.
- Equipment leakage is ≤100 µA in normal condition and ≤300 µA in single-fault condition.
- Applied-part leakage meets the applicable manufacturer and facility limit.
- No unresolved technical INOP, diagnostic failure, or safety defect remains.
Remove From Service
Remove the monitor from service when:
- Any required test fails.
- Protective-earth continuity fails.
- Leakage current exceeds the applicable limit.
- The housing, power cord, battery, mount, connector, or display presents a safety risk.
- The monitor fails to boot or repeatedly restarts.
- Alarm audio or visual indication is unreliable.
- Battery transfer causes shutdown.
- A required parameter is outside tolerance and cannot be calibrated or corrected.
- Liquid ingress, impact damage, electrical surge, fire, or contamination requires further inspection.
- The monitor cannot be restored to the approved clinical configuration.
The service guide directs testing to be discontinued immediately following a failed test until the device is repaired or identified as defective.
Copy/Paste CMMS Work Order Procedure
Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.