Philips PageWriter TC70

Preventive Maintenance Procedure

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

Philips

Model

PageWriter TC70

Manual Reference

Philips PageWriter TC70 Cardiograph Service Manual, publication number 453564106431, Edition 1, October 2008, Software Revision A.01.00 and higher.

Manufacturer Recommendation / AEM Status

Manufacturer-listed PM items

The manufacturer-listed care and maintenance items include:

Manufacturer-listed frequencies

From the maintenance frequency table:

Manufacturer-listed part numbers

Post-repair safety inspection requirements

The manual separates preventive maintenance from service events. It states that installation and preventive maintenance are not considered service events because the cardiograph is designed to be installed and maintained by the user.

Post-repair testing depends on the repair type:

AEM note

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Identify the Device

Action: Confirm manufacturer, model, asset ID, serial number, location, and accessories.

Expected outcome: Device identity matches the CMMS record.

Why it matters: TC70 units may have different PIM, network, barcode, cart, or power accessory configurations.

2. Remove From Active Patient Use

Action: Confirm the cardiograph is not connected to a patient. Power down or place in Standby before inspection and cleaning.

Expected outcome: PM is performed only when the unit is not being used for patient care.

3. Visual Inspection

Action: Inspect the cardiograph exterior and accessories for:

Expected outcome: No apparent wear, damage, corrosion, missing hardware, or unsafe condition.

Remove from service if: AC cord damage, exposed conductors, cracked patient cable insulation, damaged PIM, liquid intrusion, unstable cart mounting, damaged touch screen, or failed visual inspection is found.

4. Cleaning Check

Action: Clean external cardiograph surfaces and PIM using a soft cloth dampened with an approved cleaning solution. Avoid wetting lead wire connectors and patient data cable connectors.

Expected outcome: External surfaces are clean, dry, and free of residue.

Manual-supported cleaners include:

Do not use: strong solvents, abrasive cleaners, acetone, iodine-based cleaners, phenol-based cleaners, ethylene oxide sterilization, autoclave, ultrasonic cleaning, or immersion.

5. Patient Data Cable and Lead Wire Inspection / Cleaning

Action: Inspect patient data cable and lead wires for cracked insulation, exposed conductors, damaged connectors, loose lead adapters, or contamination. Clean with approved solution using a damp cloth.

Expected outcome: Cable and lead wires are intact, clean, dry, and connectors are not wet.

Remove from service if: cable integrity is questionable or exposed conductors are visible.

6. Reusable Electrode Check, If Present

Action: If reusable electrodes are stored with the unit, inspect for corrosion, residue, damage, and cleanliness.

Expected outcome: Reusable electrodes are clean, disinfected, dry, and usable.

Manual-listed frequency: reusable electrodes are cleaned after each patient use.

Note: Philips recommends disposable electrodes for patient applications. Welsh bulb electrodes have special warnings and should not be treated like routine disposable ECG electrodes.

7. Printer Paper and Paper Path Check

Action: Open the paper drawer. Confirm Philips-approved paper is loaded correctly with the paper sensor hole properly positioned. Replace paper if the red stripe is present or paper is low.

Expected outcome: Paper is loaded correctly, drawer closes fully, and paper advances properly.

Remove from service if: paper tray is damaged, paper will not advance, printer door/paper jam errors persist, or print quality cannot be corrected.

8. Print Head Cleaning, If Needed

Action: If print quality is uneven, open the paper drawer and lightly wipe the print head with a foam swab dipped in 90% alcohol. Allow the print head to dry before use.

Expected outcome: Print quality improves and no residue remains.

Manual-listed frequency: when necessary due to uneven print quality.

9. Battery Inspection and Status Check

Action: Confirm both batteries are installed. Check battery status from the status bar / battery status window. Confirm the unit charges on AC power and the AC power indicator illuminates.

Expected outcome: Both batteries are present, charging/charged status is reasonable, and battery condition supports mobile use.

Manufacturer-listed battery maintenance points:

10. Battery Calibration, If Needed

Action: If displayed battery status is inaccurate or not useful, perform battery calibration per the manual.

Expected outcome: Battery status information becomes more accurate.

Manual-listed frequency: when necessary due to inaccurate battery status information.

PM note: The manual states the battery calibration procedure may take up to 8 hours with the cardiograph out of active use. Philips also lists an external battery charger/calibrator, part number 989803162021.

11. Touch Screen Check

Action: Verify touch screen response across normal use areas. Clean the touch screen with a damp soft cloth using water or isopropyl alcohol. Allow it to dry.

Expected outcome: Touch screen responds accurately without repeated presses.

Manual-listed frequency:

If needed: Run touch screen calibration using Ctrl + Alt + C or the Service Utilities procedure.

12. Power On Test

Action: Restart the cardiograph.

Expected outcome: Software identification screen appears, audible beep occurs, and the Main screen loads without errors.

Self-test areas listed by the manual include:

Remove from service if: device does not power on, does not complete startup, shows persistent errors, or is unresponsive.

13. Functional Check

Action: Perform basic operational checks:

Expected outcome: Controls, display, printer, PIM, and lead system operate normally.

14. Cardiograph Overall Sensitivity Test

Action: Perform the manufacturer-listed yearly sensitivity test using a calibrated ECG simulator.

Procedure summary:

Expected outcome: The 1 mV square wave measures 10 mm with a ±5% margin for error.

Pass range: approximately 9.5 mm to 10.5 mm.

If failed: Repeat the test. If it fails again, test the simulator on another cardiograph. If the simulator passes on another cardiograph, remove this cardiograph from service and escalate.

Electrical Safety Testing if applicable

Manufacturer-listed electrical safety context

The manual identifies the TC70 as an electrical Class II device. For this device, protection against electric shock does not rely on a protective earth conductor.

The manual does not list routine annual PM electrical safety testing as a manufacturer-listed PM frequency. It does list safety tests for post-service verification, especially after internal repairs.

Ground resistance

Protective earth / ground resistance testing is not manufacturer-listed as a routine PM item for this Class II cardiograph.

Document as:

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:

Result = N/A - Class II device / no manufacturer-listed protective earth resistance PM test

Pass / Fail = N/A

If facility policy requires a ground continuity-style check on the cart, power cord, or system configuration, document the facility limit used.

Leakage current

Use facility policy and the applicable electrical safety analyzer procedure unless the repair event requires the manual’s S2/S3 safety test blocks.

Manufacturer-listed post-repair safety test limits include:

Default PM documentation:

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Applied part / patient lead leakage, if performed:

Facility limit used =

Result =

Pass / Fail =

Pass / Fail Criteria

PM pass criteria

Pass the PM if:

Remove from service criteria

Remove from service if:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide