Siemens Healthineers CLINITEK Status+

Preventive Maintenance Procedure

On this page

Asset Type

Urine Analyzer

Manufacturer

Siemens Healthineers

Model

CLINITEK Status+

Manual Reference

Bayer CLINITEK Status Operation / Service Manual

Manufacturer Recommendation / AEM Status

Manufacturer-listed PM items

The provided manual lists preventive maintenance around cleaning, inspection, and routine checks of the CLINITEK Status analyzer.

Manufacturer-listed items include:

Manufacturer-listed frequencies

Manufacturer-listed parts / materials

Post-repair safety / release requirements

The manual lists release testing after service, including:

AEM note

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Confirm Device Status

Action: Confirm the analyzer is not actively being used for patient testing. Notify the department that PM is being performed.

Expected outcome: No active urine strip or cassette test is interrupted.

2. External Visual Inspection

Action: Inspect the analyzer housing, touchscreen, printer cover, power connection, RS-232 connector, battery compartment, and test table opening. Look for cracks, missing parts, fluid intrusion, dried urine, damaged connectors, or printer compartment contamination.

Expected outcome: Analyzer is physically intact, cleanable, and safe to handle.

3. Remove and Inspect Test Table / Insert

Action: With power on, gently pull the test table carriage from the instrument. Remove the test table insert and drain the drip tray if required.

Expected outcome: Test table and insert are removable and not physically damaged.

4. Inspect Calibration Strip

Action: Examine the white calibration strip for dust, dirt, scratches, or mars. Do not touch the calibration strip.

Expected outcome: Calibration strip is clean and undamaged.

Fail condition: If the calibration strip cannot be cleaned, is scratched, or is marred, replace the test table before returning the analyzer to service.

5. Clean Test Table Insert and Drip Tray

Action: Using a cotton-tipped applicator and distilled water, scrub the strip channel, surrounding area, and drip tray. Rinse the test table insert under running water. Dry with a soft lint-free cloth or tissue.

Expected outcome: Insert, strip channel, and drip tray are clean and dry.

6. Clean Calibration Strip If Needed

Action: If the calibration strip appears dirty, gently clean it with a new cotton-tipped applicator and distilled water. Allow it to air dry and inspect again.

Expected outcome: Calibration strip is clean, dry, and undamaged.

Fail condition: Replace the test table if the calibration strip cannot be cleaned or shows scratches/marring.

7. Disinfect Test Table and Insert If Needed

Action: If disinfection is required, use only a manufacturer-listed disinfectant. Do not allow disinfectant to contact the white calibration strip. Soak the table or insert for no longer than 10 minutes, rinse thoroughly with water, dry, and reinstall.

Expected outcome: Table and insert are disinfected without damaging the calibration strip.

8. Remove Heavy Buildup If Present

Action: If dried urine buildup is present, wet a cotton-tipped applicator with 0.1N sodium hydroxide and clean the strip holding channel, surrounding areas, drip tray, and table insert. Do not allow sodium hydroxide to contact the white calibration strip. Rinse thoroughly and dry.

Expected outcome: Heavy buildup is removed without damage to the calibration strip.

9. Clean Exterior

Action: Turn the instrument off by pressing the power button for 2 seconds. Wipe the exterior using a damp cloth with mild detergent or a listed disinfectant. Do not allow liquid into the printer compartment.

Expected outcome: Exterior is clean, dry, and free of visible urine, dust, or dirt.

10. Reinstall Test Table and Insert

Action: Reinstall the test table and insert by gently pushing the table into the instrument until halfway in.

Expected outcome: Table inserts smoothly and is properly seated.

11. Functional Check

Action: Power the analyzer on and confirm it reaches the ready/select screen. Confirm the touchscreen responds, display is readable, printer has paper if used, and test table movement appears normal.

Expected outcome: Analyzer powers on normally and basic user-interface functions operate.

12. Optional Diagnostic Checks

Action: If facility PM policy includes diagnostics, or if there was a related complaint, run applicable diagnostics from the instrument diagnostic menu:

Expected outcome: Applicable diagnostics pass.

Fail condition: Failed diagnostics require troubleshooting, repair, or removal from service.

Electrical Safety Testing if applicable

The provided manual lists line leakage current as 170 microamperes in normal condition and states that the power supply is double insulated with no leakage in fault condition. The manual does not list a routine PM electrical safety test interval.

Use facility policy for electrical safety testing frequency and limits.

Ground resistance

Not applicable unless facility policy identifies a protective earth test point. The power supply is listed as double insulated.

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different.

Result = N/A / measured result =

Touch/chassis leakage

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Power supply line leakage, if tested

Manual-listed normal condition reference = 170 µA

Result =

Pass / Fail =

Pass / Fail Criteria

PM passes when:

Remove from service when:

Copy/Paste CMMS Work Order Procedure

Optional Short Procedure Block - Scheduled PM

Optional Short Procedure Block - Post-Repair Safety / Release Inspection

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide