Stryker InTouch

Preventive Maintenance Procedure

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

InTouch

Manual Reference

InTouch Critical Care Bed Maintenance Manual, 2141-009-002 REV A, dated 2012/06.

Manufacturer Recommendation / AEM Status

Stryker recommends checking the preventive maintenance items annually and states that preventive maintenance should be performed at least once a year to ensure all bed features function properly.

Manufacturer-Listed Battery Replacement

Replace both main batteries every two years.

Main battery part number: QDF9188.

Replace both batteries at the same time.

Replace the batteries immediately if the terminals are corroded, the cases are cracked, expanded, or bulging, or the batteries no longer maintain a full charge.

Manufacturer-Listed Electrical Safety Limits

Ground impedance: Not more than 100 milliohms, equivalent to 0.10 Ω.

Current leakage: Not more than 300 microamps.

Calibration and AEM Note

The manual does not list a separate recurring interval for scale calibration, bed calibration, or touchscreen calibration. The annual checklist requires verification that these systems are accurate and operating properly.

If a facility performs this PM at a different interval or changes the listed activities, the changes should be managed through the facility’s approved alternative equipment maintenance program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

The manual specifically uses a digital level or inclinometer during bed calibration. Do not zero or calibrate the inclinometer after placing it on the bed frame during the calibration procedure.

Step-by-Step PM Procedure

1. Identify and Prepare the Bed

Action:

Expected outcome: The bed is correctly identified, stable, powered, and safe for inspection.

2. Inspect the Frame and Hardware

Action:

Expected outcome: No structural damage, loose hardware, or damaged boards are present.

Remove from service when: Structural damage, cracked load-bearing components, loose critical hardware, or unsafe deformation is found.

3. Inspect the Mattress and Retainers

Action:

Expected outcome: The mattress surface is intact and remains properly secured to the bed.

Remove from service when: The mattress cover is damaged in a way that may allow fluid entry.

4. Check Casters, Brakes, and Steering

Action:

Expected outcome: The bed remains stationary with the brakes applied and rolls and steers normally with the brakes released.

5. Inspect Ground Chains

Action:

Expected outcome: Ground chains are clean, intact, securely mounted, and contact the floor.

6. Inspect Siderails

Action:

Expected outcome: Each siderail moves normally and positively locks without excessive play.

Remove from service when: A siderail does not lock securely, releases unexpectedly, or has structural damage.

7. Check CPR Operation

Action:

Expected outcome: The emergency CPR release lowers the bed correctly and normal powered operation returns afterward.

8. Check Bed Motion Functions

Action:

Operate the bed through its normal range using the siderail and footboard controls:

Listen for abnormal noise and watch for binding, hesitation, unintended movement, cable interference, or motor-overload messages.

Expected outcome: All sections move smoothly, stop correctly, and respond from each applicable control panel.

9. Check Control Panels and Indicators

Action:

Expected outcome: Controls respond consistently, indicators match the selected bed state, and the touchscreen is accurately aligned.

10. Check Scale and Bed Exit Systems

Action:

Expected outcome: The scale and Bed Exit systems operate normally without errors.

Manufacturer-listed scale accuracy:

11. Verify Angle Accuracy

Use a digital level or inclinometer to compare the actual bed-section angle with the displayed angle.

Manufacturer-listed ranges:

Action:

Expected outcome: Displayed angles correspond with the physical position and remain stable throughout bed-height movement.

Calibration mode is not necessary during every PM when the readings are accurate. Perform bed calibration only when verification fails or after applicable repairs.

12. Check Main Batteries

Action:

Expected outcome: The batteries are within the two-year interval, undamaged, securely connected, and able to support normal battery-powered operation.

Replacement battery: QDF9188.

Replacement requirement: Replace both batteries together.

13. Check Power Cord, Wiring, and Grounding

Action:

Expected outcome: The power cord, wiring, connections, and grounding hardware are intact and secure.

Remove from service when: The power cord is damaged or any exposed conductor, loose protective-earth connection, burned connection, or shock hazard is found.

14. Check Optional Zoom Drive — Model 2141

Action:

Expected outcome: The Zoom drive engages, moves, releases, and returns to neutral correctly.

15. Check Optional iBed Awareness Functions

Action:

Expected outcome: The iBed status display and light bars correctly indicate the physical bed configuration.

16. Review Maintenance Diagnostics

Action:

Expected outcome: No active or unresolved condition remains that could affect safe operation.

17. Final Functional Check

Action:

Expected outcome: The bed is complete, clean, correctly assembled, and ready for clinical use.

Electrical Safety Testing

The InTouch is Class I, Type B equipment. Perform testing with the bed connected and configured according to the electrical safety analyzer instructions and facility procedure.

Protective Earth / Ground Impedance

Manufacturer-listed limit: Not more than 100 milliohms, equivalent to 0.10 Ω.

Ground resistance measured in ohms, expected ≤ 0.10 Ω per the manufacturer:

Result =

Pass / Fail =

The manufacturer limit is more restrictive than the commonly used 0.5 Ω facility threshold and should be used unless an approved facility policy or AEM determination establishes otherwise.

Leakage Current

Manufacturer-listed limit: Not more than 300 microamps.

Touch/chassis leakage measured in microamps:

Facility limit used = 300 µA manufacturer limit

Result =

Pass / Fail =

Document the analyzer mode and test conditions according to facility electrical safety procedures.

Pass / Fail Criteria

Pass

The bed passes when:

Remove From Service

Remove the bed from service when:

Copy/Paste CMMS Work Order Procedure

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide