On this page
Asset Type
Manufacturer
Model
Manual Reference
LUCAS 3 Chest Compression System Service Manual, 3328798-002, ©2017 Physio-Control, Inc.
Manufacturer Recommendation / AEM Status
Manufacturer-Listed PM Items
The manufacturer-listed preventive maintenance procedure includes:
- Cleaning the device according to the Instructions for Use.
- Check/replace bellows if not intact or clean.
- Check/replace suction cup if not intact or clean.
- Check/replace patient straps.
- Check/replace stabilization strap.
- Check/replace carrying bag.
- Check/update device software.
- Check for errors and warnings by starting the device and letting it run for one minute.
- Clean electric fan and mesh grill.
- Lubricate the compression module.
- Lubricate the claw lock.
- Perform a function check.
The service manual notes to use the PIP checklist 3326789 Rev C to record PIP results during the maintenance procedure.
Manufacturer-Listed Frequencies
PM frequency: Not specified in the provided service manual.
The manual lists a preventive maintenance procedure but does not provide a time-based PM interval in the provided service manual. Facility PM interval should be based on the manufacturer IFU/PIP checklist if available, facility policy, service history, device risk, and applicable regulatory requirements.
Manufacturer-Listed Part Numbers and Substances
Manufacturer-listed substances and service materials include:
- Carry Ball Screw Grease: 21576-000018
- PTFE Spray Lubricant: 21576-000023
- Compressed Air, Can: 21300-001335
- Thread Lock Fluid: 11996-000220
Common accessory/service replacement items listed in the manual include:
- LUCAS Battery, dark gray: 11576-000080
- LUCAS 2 Suction Cup, 3-pack: 11576-000046
- LUCAS Stabilization Strap: 21576-000074
- LUCAS 2 Patient Strap pair: 11576-000050
- LUCAS 3 Refurbishing Kit: 21576-000079
- LUCAS 3 Hood Assembly: 21576-000080
- LUCAS 3 Support Leg: 21576-000081
- LUCAS 3 Control Board: 21576-000082
- LUCAS 3 Protective Board: 21576-000091
- LUCAS 3 Communication Board: 21576-000085
- LUCAS 3 Version 3.1 Communication Board: 21576-000097
Post-Repair Safety Inspection Requirements
The manual states that after each repair or maintenance, a Function Check should be performed to establish all vital functions of the device.
The manual also states that a qualified Technical Support person should inspect any device that has been dropped, damaged, or abused to verify that the device is operating within the performance standards listed in the Performance Inspection Procedures.
AEM Note
If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.
What This PM Procedure Covers
- Visual and mechanical condition checks
- Cleaning review
- Bellows, suction cup, straps, and carrying bag inspection
- Software/error/warning review
- Fan and mesh grill cleaning
- Compression module lubrication
- Claw lock lubrication
- Mechanical and electronic function check
- Electrical safety documentation fields for facility policy use
Required Tools and Equipment
Manufacturer-listed standard tools include:
- Torx screwdriver T10, T20, T25 or Torx key set
- Phillips screwdriver PH1
- Small flat screwdriver
- Flat screwdriver 5.5 x 0.8 mm
- Adjustable torque screwdriver, minimum range 0.6–4.0 Nm
- PH1 bit
- Flat 5.5 x 0.8 mm bit
- Torx T10, T20, T25 bits
- Hammer
- Mandrel cutting
- Adjustable pliers
- ESD pliers
- Protection kit
- Micro SD card reader
- USB to USB mini cable
- Non-knurled flat nose plier
- USB programming cable
- REDEL extracting tool
Special tools listed:
- Hood Holder Bracket Service Tool: 100838-00
- Mandrel, LUCAS Hinge Insertion optional tool: 300040-00
- USB Programming Cable: 160423-00
Required software listed:
- LUCAS 3 Program Loader 3.0: 21340-000846
- LUCAS 3 Program Loader 3.1: 21340-000852
- LUCAS 3 Report Generator: free download from LIFENET
- DT Express: free download from LIFENET
For a routine scheduled PM that does not involve disassembly, use only the tools and software needed for the facility’s PM scope. Do not open the device unless trained, authorized, and following the service manual.
Step-by-Step PM Procedure
1. Remove Device From Clinical Readiness Area
Action: Remove the LUCAS 3 from clinical use/ready stock for scheduled PM. Verify no patient use is in progress.
Expected outcome: Device is available for inspection, cleaning verification, functional testing, and documentation.
2. Confirm Device Identification
Action: Record model, serial number, asset number, software version if checked, and battery serial number if applicable.
Expected outcome: Device identity matches the work order and service documentation.
3. Review Service History and Damage History
Action: Check for recent drops, damage, abuse, malfunction reports, failed starts, battery warnings, compression alarms, or user complaints.
Expected outcome: Any device with drop/damage/abuse history is inspected by qualified technical support and verified against the PIP/performance standard before return to service.
4. Clean Device
Action: Clean the device according to the LUCAS 3 Instructions for Use. Do not use unapproved cleaning methods.
Expected outcome: Device is clean and suitable for inspection.
Why it matters: Dirt, fluids, and debris can hide mechanical damage, affect straps and locks, and contaminate moving assemblies.
5. Inspect Bellows
Action: Check the bellows for cleanliness, cracks, tears, looseness, deformation, or improper seating.
Expected outcome: Bellows is intact, clean, and correctly installed.
Remove from service if: Bellows is torn, damaged, contaminated beyond cleaning, loose, or not properly seated.
6. Inspect Suction Cup
Action: Check suction cup for cleanliness, cracking, deformation, wear, secure attachment, or contamination.
Expected outcome: Suction cup is intact, clean, and attached correctly.
Remove from service if: Suction cup is damaged, excessively worn, contaminated, or does not attach properly.
7. Inspect Patient Straps, Stabilization Strap, and Carrying Bag
Action: Inspect all fabrics, Velcro, buckles, stitching, strap hardware, and carrying bag condition.
Expected outcome: Patient straps, stabilization strap, and carrying bag are clean and functional. Velcro and buckles are not worn and function correctly.
Remove from service if: Straps or buckles are damaged, worn, contaminated, missing, or do not secure correctly.
8. Check Battery Condition and Age/Use Indication
Action: Install a charged battery. Review battery indicator behavior. Check for battery replacement indications or warnings.
Expected outcome: Battery is recognized and supports normal device operation.
Manual-backed replacement condition: The manual lists battery-related indications for battery used more than 200 times or more than 200 minutes with compressions each month or older than 3–4 years. It also lists battery replacement recommended for “TOO MANY USES” and “TOO OLD BATTERY” charger CPU entries.
Remove from service if: Battery is not recognized, depleted, too old, over-use warning is present, damaged, leaking, or unable to support operation.
9. Check/Update Software
Action: Connect using LUCAS 3 Program Loader when required by PM scope or facility procedure. Check software version and update if applicable per the service manual and authorized service process.
Expected outcome: Software version is documented and updated if required by facility/manufacturer service process.
Do not return to service if: Software update fails, checksum/version verification fails, or service tool reports unresolved errors/warnings.
10. Check for Errors and Warnings
Action: Start the device and let it run for approximately one minute. Use LUCAS 3 Program Loader to review device errors/warnings if required by facility procedure or if any abnormal behavior is observed.
Expected outcome: No active or stored errors/warnings that affect safety or function.
Remove from service if: Any high-priority alarm, self-test fault, internal fault, compression out-of-range alarm, battery fault, temperature fault, communication fault affecting operation, or repeated warning is present.
11. Clean Electric Fan and Mesh Grill
Action: If performing full service-level PM with hood removed, cover the Carrier Ball Screw hole with tape to keep dust out of the lubrication area. From inside the device, clean the fan and mesh grill with compressed air. Remove dust from areas requiring lubrication. Remove tape when complete.
Expected outcome: Fan and mesh grill are clean, and dust is not blown into the lubrication area.
Caution: Do not contaminate the Carry Ball Screw lubrication area.
12. Lubricate Compression Module
Action: Use only Carry Ball Screw Grease 21576-000018. With the hood off, pull down the piston and adjustment tube until the Carry Ball Screw is visible in the lubrication hole. Apply 0.5 mL of grease to the Carry Ball Screw. Move the piston manually up and down 5 times over the full range.
Expected outcome: Carry Ball Screw is lubricated and piston movement is smooth.
Remove from service if: Piston binds, moves roughly, makes unusual noise, or does not move through full travel smoothly.
13. Lubricate Claw Lock
Action: Use PTFE Spray Lubricant 21576-000023. Spray a small quantity on movable claw lock parts, including gliding surfaces of the claws, axle entry points into plastic, and locking pin. Open and lock the mechanism repeatedly to work in the lubricant.
Expected outcome: Claw locks move freely and lock/unlock correctly.
Remove from service if: Claws do not lock securely, release unexpectedly, bind, or fail the mechanical claw lock function check.
14. Perform Mechanical Function Check - Piston
Action: With the device off, pull the suction cup down and up. Verify the decompression spring moves smoothly and without unusual noise. Slowly pull down until the Carry Ball Nut reaches its lower position. Push suction cup back to top position.
Expected outcome: Piston travels smoothly through full range without unusual noise.
Note: Manual warns that pushing too fast may create stepped resistance because the electric motor is affected. Move slowly in one long stroke.
15. Perform Mechanical Function Check - Claw Lock Mechanism
Action: Check play between release ring and support legs. Pull release ring gently to unlock and open claws. Verify detectable play/rattle between ring and support leg. Press claw lock closed. When locked, press hard with two thumbs on the two claws.
Expected outcome: Claw mechanism locks securely, does not release under manual pressure, and release ring has expected play.
Fail condition: If the support ring is tight against the support leg with no play, the check has failed. If the claw mechanism unlocks during the locked pressure check, the check has failed.
16. Perform Electronics Function Check - Startup
Action: Turn on the device.
Expected outcome: Internal Function Test completes and ADJUST LED shows green.
Remove from service if: Startup test fails, warning LED activates, abnormal alarm occurs, or device does not enter ADJUST mode correctly.
17. Perform ACTIVE 30:2 Check
Action: Change mode to ACTIVE 30:2. Let the device run for approximately two minutes. Listen for unusual noise. Verify audible alert before ventilation pause and intermittent LED behavior at default settings.
Expected outcome: ACTIVE 30:2 LED shows green. Device runs without abnormal noise. Ventilation alert/pause behavior is correct for configured settings.
18. Perform PAUSE Check
Action: Change mode to PAUSE.
Expected outcome: PAUSE LED shows green.
19. Perform ACTIVE Continuous Check
Action: Change mode to ACTIVE continuous.
Expected outcome: ACTIVE continuous LED shows green. At default settings, LED blinks every ten strokes for ventilation alert.
20. Perform ADJUST / Adjustment Servo Check
Action: Change mode to ADJUST. Pull suction cup down to test adjustment servo. Push suction cup back to top position.
Expected outcome: Suction cup and adjustment servo move smoothly.
Remove from service if: Motion is rough, weak, noisy, binds, or does not return correctly.
21. Check Charging Sequence
Action: Connect charger/power supply to device.
Expected outcome: Battery LEDs show running light indicating charging sequence begins. Disconnect charger after verification.
Remove from service if: Charging sequence does not begin, power inlet is damaged, charger connection is intermittent, or battery fault occurs.
22. Test Transmit Mode / Service Tool Communication
Action: Follow service manual procedure to connect through Bluetooth using LUCAS 3 Program Loader when required. Verify communication works. If maintenance passed the Function Check, press Set PC Date to update Last Service Date. Press OFF in Program Loader to switch off Bluetooth communication and verify Bluetooth light is off.
Expected outcome: Device communicates with Program Loader. Last Service Date is updated after maintenance. Bluetooth turns off properly.
Important: If the device underwent repair without maintenance, the manual says not to press Set PC Date.
23. Final Review
Action: Confirm all checks passed, no unresolved alarms/errors/warnings remain, accessories are present, device is clean, battery is charged/acceptable, and all documentation is complete.
Expected outcome: Device is ready for return to service.
Electrical Safety Testing if Applicable
The provided service manual does not list a routine electrical safety test requirement or leakage current limit for scheduled PM.
Perform electrical safety testing if required by facility policy, after applicable repairs, after power inlet/power supply concerns, or under the facility’s electrical safety program.
Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:
Result =
Touch/chassis leakage measured in microamps:
Facility limit used =
Result =
Pass / Fail =
Pass / Fail Criteria
PM Pass Criteria
Device may pass PM when:
- Device is clean.
- Bellows is intact, clean, and correctly installed.
- Suction cup is intact, clean, and correctly installed.
- Patient straps, stabilization strap, and carrying bag are clean and functional.
- Battery is accepted by the device and has no replacement warning/fault condition.
- Software/error/warning review is acceptable.
- Fan and mesh grill are clean when service-level PM is performed.
- Compression module lubrication is completed with specified grease when service-level PM is performed.
- Claw lock lubrication is completed with specified PTFE lubricant when service-level PM is performed.
- Mechanical piston check passes.
- Claw lock mechanism check passes.
- Electronics function check passes.
- Charging sequence check passes.
- Program Loader communication/transmit check passes when performed.
- No unresolved alarms, errors, warnings, or abnormal noises remain.
- Electrical safety testing passes if performed by facility policy.
Remove From Service Criteria
Remove the LUCAS 3 from service if any of the following are found:
- Failed internal function test.
- High-priority alarm.
- Self-test fault.
- Internal fault.
- Compression out-of-range condition.
- Disruption of suction cup alarm that cannot be resolved.
- Battery depleted, too old, too many uses, not recognized, damaged, leaking, or unable to support operation.
- Bellows or suction cup damaged.
- Straps, buckles, Velcro, stabilization strap, or carrying bag damaged or nonfunctional.
- Claw lock fails to lock securely or releases during locked check.
- Piston movement is rough, noisy, binding, or incomplete.
- Charging sequence does not start.
- Program Loader reports unresolved safety/function errors.
- Device has been dropped, damaged, or abused and has not been verified by qualified technical support against the PIP/performance standards.
- Electrical safety test fails when performed.
Copy/Paste CMMS Work Order Procedure
Scheduled PM Short Procedure
Post-Repair Function Check
Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.