Stryker ProCuity

Preventive Maintenance Procedure

On this page

Asset Type

Hospital Bed

Manufacturer

Stryker

Model

ProCuity

Manual Reference

ProCuity Bed Series Maintenance Manual, 3009-309-002 Rev AD.0, October 2025

Manufacturer Recommendation / AEM Status

Stryker lists preventive maintenance annually for the ProCuity Bed Series. The manual instructs the technician to remove the bed from service before inspection and states that more frequent maintenance may be necessary based on product usage.

Manufacturer-Listed PM Activities

Manufacturer-Listed Limits and Service Information

The annual PM checklist separately lists replacement of the bed batteries. The difference between the two-year expected battery service life and annual battery replacement checklist should be reviewed with Stryker or addressed through approved facility policy rather than silently changing the checklist.

Service procedures throughout the manual generally instruct the technician to verify proper operation before returning the product to service.

If the facility changes the annual interval or modifies the manufacturer-listed activities, document the change under the facility’s approved Alternative Equipment Maintenance program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

The maintenance manual specifically identifies an inclinometer for angle calibration when the main control board has been replaced. Routine annual Fowler-angle verification may use the bed indication and facility-approved measurement equipment.

Step-by-Step PM Procedure

1. Remove the Bed From Service

Action

Expected Outcome

The bed is clean, safely isolated, and ready for inspection.

2. Inspect the Frame and Hardware

Action

Inspect for:

Tighten accessible loose hardware when appropriate. Escalate damaged structural components for repair.

Expected Outcome

All hardware is secure, and no structural damage, fluid leakage, or unsafe condition is present.

3. Inspect the Mattress and Support Surface

Action

Inspect the mattress cover for:

Inspect powered mattress connections when installed.

Expected Outcome

The support surface is intact and suitable for clinical use.

4. Test Brakes and Casters

Action

Expected Outcome

The bed remains stationary with brakes applied, rolls normally when released, and all brake and steer indicators operate correctly.

5. Inspect and Test Siderails

Action

Expected Outcome

All siderails move smoothly, latch securely, and all controls and indicators respond correctly.

6. Test CPR Release

Action

Expected Outcome

Both CPR releases operate correctly, and normal Fowler operation is restored after testing.

7. Test Bed Motions

Action

Operate the full available range of:

Confirm that the motion interrupt below the bed extender operates.

Expected Outcome

All motions are smooth, correctly directed, and stop when released. The motion-interrupt system responds correctly.

8. Verify Fowler Angle

Action

Expected Outcome

The Fowler angle indication corresponds with actual Fowler position throughout the listed range.

9. Calibrate the Scale and Bed Exit System

Action

Expected Outcome

Scale readings are accurate and stable, and the Bed Exit system responds correctly.

Scale calibration result:
Applied test weight =
Displayed weight =
Difference =
Pass / Fail =

10. Test Controls, Indicators, and Night Light

Action

Test:

Expected Outcome

All controls respond correctly, displays are readable, and indicators operate as intended.

11. Test Nurse Call

Action

Expected Outcome

The nurse-call signal is transmitted and cleared correctly.

Nurse-call result:
Pass / Fail =

12. Inspect Power Cords, Cables, and Connections

Action

Inspect:

Confirm that all grounding connections are secure to the frame.

Expected Outcome

No cord, cable, connector, or grounding defect is present.

13. Inspect the Ground Chain

Action

Inspect the ground chain for cleanliness and damage.

Confirm:

Expected Outcome

The ground chain is clean, secure, and maintains appropriate floor contact.

14. Replace the Main Bed Batteries

Action

The annual PM checklist lists replacement of the bed batteries.

Expected Outcome

New batteries are securely installed, battery statistics are reset, and the bed operates normally on battery power.

Battery replacement note: The manual’s expected-service-life section states two years for backup batteries, while its annual checklist lists battery replacement. Document the facility’s resolution of this discrepancy.

Battery replacement completed =
Battery date code =
Battery statistics reset = Yes / No

15. Test Applicable Options

Perform only those checks applicable to the bed configuration.

iBed Wireless / IR Module

Zoom Drive

IV Pole

Electric Brakes and Secure Connect

Expected Outcome

Every installed option operates correctly and shows no physical damage.

16. Final Functional Inspection

Action

Expected Outcome

The bed is fully assembled, calibrated, electrically safe, and ready for service.

Electrical Safety Testing

The ProCuity maintenance manual provides specific electrical safety limits. Document the manufacturer-listed values rather than substituting the general 0.5 Ω ground-resistance criterion used when no device-specific value is available.

Ground Impedance

Manufacturer limit: Not more than 200 mΩ, equivalent to 0.200 Ω.

Ground resistance measured in ohms, expected ≤ 0.200 Ω:
Result =
Pass / Fail =

Leakage Current

Manufacturer limit: Not more than 300 µA.

Touch/chassis leakage measured in microamps:
Facility limit used = 300 µA
Result =
Pass / Fail =

Document the electrical-safety analyzer configuration, test conditions, and applied-part classification according to facility policy.

Pass / Fail Criteria

Pass

The bed passes PM when:

Remove From Service

Remove the bed from service for:

Copy/Paste CMMS Work Order Procedure

Scheduled PM

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?