Tuttnauer 3870

Preventive Maintenance Procedure

On this page

Asset Type

Sterilizer

Manufacturer

Tuttnauer

Model

3870

Manual Reference

Operation & Maintenance Manual, Tabletop Autoclaves Models 1730, 2340, 2540, 3140, 3850, 3870 M & MK, Cat. No. MAN205-0007000EN, Rev. Q.

Manufacturer Recommendation / AEM Status

Tuttnauer lists maintenance activities at daily, weekly, monthly, 12-month, annual, and cycle-based intervals.

Daily

Clean the door gasket using mild detergent, water, and a soft cloth or sponge. Confirm the gasket is clean and smooth. Before operation, check the reservoir water level and add distilled water when needed.

Weekly

Clean the air jet. Clean and descale the chamber, copper tubes, and reservoir using Chamber Brite. Remove and clean the trays and tray holder with detergent or a non-abrasive stainless-steel cleaner. Lubricate the two door pins, door-tightening bolt screw shaft, and bearing. Clean the exterior with a soft cloth.

Weekly or Every 20 Cycles, Whichever Occurs First

Replace the water in the reservoir.

Monthly

Clean and check the safety valve. The manual recommends having the distilled or mineral-free water quality tested once per month.

Every 12 Months or as Needed

Replace the door gasket.

Annually

Inspect the locking device for excessive wear.

Chamber Brite Interval Clarification

Section 8.1 lists Chamber Brite cleaning once per week. The detailed Chamber Brite procedure also states every 20 cycles or as needed. Because both intervals appear in the manual, follow the more frequent interval or the facility’s documented maintenance policy.

The 3870 requires two packets of Chamber Brite per cleaning procedure. The listed cleaning cycle for an M-series autoclave is 30 minutes at 273°F/134°C from a cold start, without a drying cycle.

Manufacturer-Listed Replacement Items

Door gasket: Replace every 12 months or as needed.

Door gasket part number: Not specified in the provided manual.

Electrical Safety and Calibration Status

The provided operation and maintenance manual does not list a recurring electrical safety testing interval, ground resistance or leakage-current limits, a pressure-gauge calibration interval, a thermostat calibration interval, or a manufacturer PM test form with calibration tolerances.

The unit is intended to be connected to a properly grounded outlet protected by an earth-leakage device or GFI. Electrical safety testing should be completed and documented when required by facility policy.

If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

Step-by-Step PM Procedure

1. Remove the Sterilizer from Service

Action

Expected Outcome

The unit is electrically isolated, depressurized, and cool enough to service safely.

2. Inspect the Exterior and Installation

Action

Inspect the cabinet and control panel, power cord, plug, rear power inlet, main switch, pressure gauge, timer and thermostat knobs, multipurpose valve and knob, indicator lights, feet, countertop stability, and rear and right-side ventilation clearance.

Confirm the sterilizer is level and positioned on a rigid surface capable of supporting the unit and load.

Expected Outcome

No damaged, loose, burned, cracked, corroded, or missing components are present.

3. Inspect and Clean the Door Gasket

Action

Expected Outcome

The gasket is clean, flexible, undamaged, correctly installed, and capable of sealing the chamber.

4. Inspect the Door and Locking Mechanism

Action

Inspect the door closing device, tightening bolt and threads, two door pins, door microswitch and actuator, pressure-locking bellows mechanism, door alignment, and locking components for excessive wear.

Apply a few drops of oil to the two door pins, door-tightening bolt screw shaft, and closing-device bearing.

Do not overtighten the door. The manual instructs that it be tightened only hand-tight unless additional tightening is needed to obtain a complete seal.

Expected Outcome

The door closes smoothly, activates the microswitch, locks securely under pressure, and shows no excessive wear.

5. Remove and Clean the Trays and Tray Holder

Action

The manual lists the following 3870 tray components:

Expected Outcome

Trays and holder are clean, undamaged, and securely supported.

6. Drain and Flush the Reservoir

Action

Do not fill above the base of the safety-valve holder.

Expected Outcome

The reservoir drains freely, the drain valve seals, and the reservoir contains clean distilled water at the correct level.

7. Clean the Air Jet

Action

The manual identifies a dirty air jet as a primary cause of incomplete air removal, failed spore tests, and chemical indicators that do not change correctly.

Expected Outcome

The air-jet opening is clear and the cleanout wire moves freely.

8. Clean and Descale with Chamber Brite

Action

Do not process instruments during the cleaning procedure.

Do not use bleach, steel wool, steel brushes, or abrasive cleaning materials.

Expected Outcome

The chamber, reservoir, copper tubing, and associated piping are clean and free of visible deposits and Chamber Brite residue.

9. Check the Chamber Pitch and Fill Level

Action

Expected Outcome

The chamber pitch allows the correct quantity of water to cover the chamber bottom without excessive or insufficient fill.

10. Perform the Monthly Safety-Valve Check

Action

Use heat-resistant PPE and keep your face and hands away from escaping steam.

If the valve remains stuck open, allow pressure to fall to zero, operate the valve ring, and repeat the check. If it sticks open again, remove the sterilizer from service and arrange repair.

Expected Outcome

The valve releases steam when operated and reseats immediately without continued leakage.

11. Perform a Functional Cycle

Action

For a 3870 M cold-start cycle, the manual lists:

These are total sterilization-cycle timer settings and do not include drying.

Expected Outcome

The sterilizer heats, pressurizes, controls, times, exhausts, dries, and releases the door normally without steam or water leakage.

12. Verify Sterilization Monitoring

Action

When required by facility infection-prevention policy:

The manual states that a biological spore test should be used at least weekly, while noting that testing standards may vary.

Expected Outcome

Sterilization-monitoring results comply with the facility’s approved policy.

13. Final Inspection

Action

Inspect for door-gasket leakage, safety-valve leakage, steam escaping from the closing-device area, water leakage from the drain valve, water spilling when the door opens, abnormal noise, odor, overheating, or electrical operation, failure of pressure to rise or return to zero, and failure of the door interlock.

Reinstall the trays and tray holder after all cleaning residue has been removed.

Expected Outcome

The sterilizer is clean, complete, leak-free, and operating normally.

Electrical Safety Testing

The provided manual does not state a recurring electrical safety test interval or manufacturer-specific acceptance limits.

Perform electrical safety testing when required by facility policy, after applicable electrical repairs, or under the facility’s approved AEM program.

Ground Resistance

Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different:

Result =

Leakage Current

Touch/chassis leakage measured in microamps:

Facility limit used =

Result =

Pass / Fail =

Additional Electrical Inspection

Protective ground present at the supply outlet: Pass / Fail

Power cord and plug condition: Pass / Fail

GFI or earth-leakage protection present as required by installation policy: Pass / Fail

Pass / Fail Criteria

Pass

The sterilizer passes PM when:

Remove from Service

Remove the sterilizer from service when:

Copy/Paste CMMS Work Order Procedure

Chamber Brite Cleaning

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide