ZOLL AED Plus

Preventive Maintenance Procedure

On this page

Asset Type

Defibrillator

Manufacturer

ZOLL

Model

AED Plus

Manual Reference

AED Plus Administrator’s Guide, REF 9650-0301-01 Rev. X, March 2016

Manufacturer Recommendation / AEM Status

The manufacturer does not provide a single fixed annual or semiannual preventive-maintenance interval in the provided manual.

Manufacturer-Listed Maintenance Direction

Manufacturer-Listed Battery Interval

Manufacturer-Listed Parts and Accessories

AEM Status

The facility should establish the scheduled PM frequency through its maintenance policy because the manual directs the user to inspect the device frequently, as necessary, but does not assign a fixed external PM interval.

If the facility performs this PM at a different interval or changes the listed activities, the changes should be managed through the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.

What This PM Procedure Covers

Required Tools and Equipment

A simulator alone confirms functional shock delivery but does not verify delivered-energy accuracy. Use a defibrillator analyzer when energy accuracy is part of the facility PM procedure.

Step-by-Step PM Procedure

1. Remove the AED from Active Readiness

Action: Remove the AED Plus from clinical readiness and provide another verified AED for coverage during testing.

Expected outcome: Emergency-response coverage remains available throughout the PM.

2. Review Device History and Configuration

Action: Record the following information:

Hold the On/Off button for at least five seconds during the manual self-test to display the software version on units equipped with an LCD.

Expected outcome: The correct battery-replacement interval and applicable tests can be determined.

3. Inspect the Exterior

Action: Inspect the enclosure, cover, handle, labels, battery door, status window, controls, display, speaker openings, electrode connector, and IrDA area.

Check for:

Expected outcome: The device is clean, structurally intact, and free of damage that could affect safe operation.

4. Inspect the Electrode Package and Cable

Action: Verify that:

Replace expired, opened, damaged, or used electrodes.

Expected outcome: A compatible, sealed, in-date electrode package is connected and ready for use.

5. Inspect Batteries and Determine the Replacement Due Date

Action: Check:

Apply the manufacturer-listed replacement interval:

Expected outcome: All batteries are within the applicable manufacturer-listed replacement period and show no evidence of leakage, corrosion, swelling, or damage.

Battery Replacement, When Due

  1. Turn the AED off.
  2. Remove the battery-compartment cover.
  3. Remove all ten batteries.
  4. Install ten new compatible Type 123A lithium manganese dioxide batteries with correct polarity.
  5. Confirm that all batteries are securely seated.
  6. Press the battery reset button when prompted.
  7. Do not press the reset button unless all ten batteries are new.
  8. Install the battery cover securely.
  9. Document the battery date and next replacement due date.

Expected outcome: All ten batteries are new, correctly installed, and the battery-use indicator has been properly reset.

6. Verify Standby Readiness

Action: With the device off, inspect the status indicator.

Expected outcome:

A red X or recurring beep indicates that the AED is not ready for service.

7. Perform the Manual Self-Test

Action: Press and hold the On/Off button for approximately five seconds.

Observe and verify:

The manual self-test checks battery capacity, electrode connection, ECG circuitry, charge/discharge circuitry, processor operation, CPR circuitry, audio, and display functions.

Expected outcome: The self-test completes successfully with UNIT OK and normal visual and audio output.

8. Verify Power-Off Readiness

Action: Turn the AED off and wait approximately two minutes.

Expected outcome:

9. Perform the Optional Technical Functional Checkout

Perform this section when included by facility policy.

Connect the Tester

Action: Connect the AED Plus Simulator/Tester or equivalent to the electrode connector. Power on the simulator and AED.

Expected outcome:

Verify Shockable-Rhythm Analysis

Action: Apply a simulated ventricular-fibrillation rhythm.

Expected outcome:

Verify Shock Delivery

Action: Press the flashing Shock button.

Expected outcome:

Verify Nonshockable-Rhythm Analysis

Action: After the CPR period, or using the tester’s normal sequence, select normal sinus rhythm.

Expected outcome:

10. Verify Delivered Energy

Perform this step only when energy verification is included in the facility PM procedure.

Action: Use a defibrillator analyzer to measure delivered energy at the device’s configured energy sequence.

The factory-default sequence listed in the manual is:

Adult Mode

Pediatric Mode

The device may be configured differently. Verify the actual programmed sequence before applying pass/fail criteria.

Manufacturer-listed accuracy: ±15%

Energy Documentation

Configured adult sequence:

Shock 1 setting =
Measured energy =
Expected tolerance = ±15%
Pass / Fail =

Shock 2 setting =
Measured energy =
Expected tolerance = ±15%
Pass / Fail =

Shock 3 setting =
Measured energy =
Expected tolerance = ±15%
Pass / Fail =

Configured pediatric sequence, when tested:

Shock 1 setting =
Measured energy =
Expected tolerance = ±15%
Pass / Fail =

Shock 2 setting =
Measured energy =
Expected tolerance = ±15%
Pass / Fail =

Shock 3 setting =
Measured energy =
Expected tolerance = ±15%
Pass / Fail =

Expected outcome: Each measured value is within ±15% of its configured energy setting.

11. Verify CPR Feedback

Action: Using a compatible tester with CPR simulation:

Expected outcome: CPR sensing, metronome operation, depth prompts, and CPR-cycle timing function correctly.

12. Verify Controls, Indicators, and Audio

Action: Confirm operation of:

Expected outcome: All controls, indicators, displays, and prompts operate clearly and consistently.

13. Verify Rescue Supplies

Action: Confirm required supplies are available according to facility configuration, such as:

Expected outcome: The AED kit is complete and ready for emergency use.

14. Clean the Device

Action: Clean the AED Plus with a soft, damp cloth using one of the manufacturer-listed methods:

Do not:

Expected outcome: The device is clean and dry without damage to the enclosure, labels, display, or connectors.

15. Perform the Final Readiness Check

Action: Reconnect and correctly store the electrode package. Close the cover. Turn the device off and allow the standby indicator to stabilize.

Expected outcome:

Electrical Safety Testing

The provided manual does not list a routine ground-resistance test, leakage-current test, or electrical-safety interval for the AED Plus.

The device is internally powered and identified as Class II equipment. Do not apply a protective-earth resistance requirement to this device.

Electrical safety testing, when performed, should follow facility policy and the applicable test method for an internally powered, defibrillation-protected Type BF device.

Facility-Policy Documentation

Electrical safety testing performed: Yes / No / Not applicable

Test standard or facility procedure used =

Touch/chassis leakage measured in microamps:
Facility limit used =
Result =
Pass / Fail =

Patient-lead or applied-part leakage, if performed:
Facility limit used =
Result =
Pass / Fail =

No manufacturer-listed leakage-current limit was found in the provided manual.

Pass / Fail Criteria

Pass

The AED Plus may pass PM when:

Remove From Service

Remove the AED Plus from service when:

Copy/Paste CMMS Work Order Procedure

Battery Change Only

Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.

Was this PM procedure helpful?

Don't see a guide you need?

Suggest a Guide