On this page
Asset Type
Manufacturer
Model
Manual Reference
ZOLL R Series Service Manual, 9650-0903-01 Rev. L, April 2016
Manufacturer Recommendation / AEM Status
The ZOLL R Series Service Manual lists two levels of checkout:
- Operator-level defibrillator testing checklist.
- More extensive 12-month maintenance test checkout procedures performed by a qualified biomedical technician.
The manual states that a qualified biomedical technician must perform the thorough maintenance test checkout every 12 months.
Manufacturer-listed 12-month maintenance tests
- Physical inspection
- Front panel button test
- 3-lead and 5-lead ECG test
- Power supply test — listed as optional
- Leakage current test
- Paddles test
- Heart rate display test
- Calibration pulse / strip chart test
- Notch filter test
- Heart rate alarm test
- Defibrillator self-test
- Synchronized cardioversion test
- Remote sync cardioversion test — optional / if applicable
- Shock test
- Summary report test
- Advisory message test — manual/advisory units
- Pacer test — if pacing option installed
- SpO₂ monitor test — if SpO₂ option installed
- EtCO₂ monitor test — if EtCO₂ option installed
- Barometric pressure calibration check — if EtCO₂ option installed
- CO₂ accuracy check — if EtCO₂ option installed
- NIBP volume leak test — if NIBP option installed
- NIBP transducer calibration test — if NIBP option installed
- NIBP monitor test — if NIBP option installed
Manufacturer-listed frequencies
- Full maintenance test checkout: Every 12 months
- NIBP volume leak test: Annually or every 10,000 readings, whichever comes first
- NIBP transducer calibration test: Annually or every 10,000 readings, whichever comes first
- Other specific interval differences: Not specified in the provided manual beyond the 12-month maintenance test structure.
Part numbers listed in maintenance section
- 3-lead adapter: 8009-0762-XX
- ZOLL Medical Electrode Adapter from Fluke Biomedical / DNI: 3010-0378
- 2-Post Battery Fixture: 9100-0575-TF
AEM note
If a facility performs this PM at a different interval or changes the listed activities, that should be handled under the facility’s approved AEM program, risk assessment, maintenance history, and applicable regulatory requirements.
Because this is a defibrillator, shortening the PM too aggressively or skipping defibrillator output, sync, leakage, pacing, or option-specific tests should be treated as a high-risk AEM decision, not a casual checklist edit.
What This PM Procedure Covers
- Annual manufacturer-listed maintenance checkout
- Visual and physical inspection
- Defibrillator output and charge performance
- Synchronized cardioversion verification
- ECG, alarm, recorder, and front panel function checks
- Leakage current documentation
- Optional pacing, SpO₂, EtCO₂, and NIBP checks when installed
- Copy/paste CMMS documentation
Required Tools and Equipment
Use equivalent test equipment only when appropriate and consistent with manufacturer and facility policy.
- Defibrillator analyzer, such as Fluke Impulse 4000 with software 1.06 or higher
- Electrical safety analyzer / leakage current tester
- ZOLL Medical Electrode Adapter / DNI 3010-0378 or equivalent
- Fully charged ZOLL SurePower defibrillator battery
- Extra fully charged battery
- AC line cord
- Paddles, if applicable
- Printer paper
- ECG cable, 3-lead adapter 8009-0762-XX, 3-lead ECG cable, and 5-lead ECG cable
- Stopwatch
- Multimeter, such as Fluke 75 or equivalent
- DC power supply, 15 A minimum, for optional power supply testing
- 0.1 Ω 1% resistor, ¼ W or greater
- 1000 Ω 1% ¼ W resistor
- Red and black miniature alligator test leads
- SpO₂ simulator, SpO₂ cable, and sensor if SpO₂ option is installed
- NIBP simulator if NIBP option is installed
- CAPNOSTAT 5 mainstream cable with airway adapter if EtCO₂ option is installed
- Gas regulator and calibration gas if performing CO₂ accuracy check
Step-by-Step PM Procedure
1. Confirm Unit Identity and Configuration
Action: Record the serial number, location, installed options, and whether the unit includes pacing, SpO₂, EtCO₂, and/or NIBP.
Expected outcome: Unit configuration is known before testing begins.
Why it matters: Several manufacturer-listed tests are option-dependent. Do not mark missing options as failed.
2. Perform Physical Inspection
Action: Inspect housing, handle, recorder door, input connectors, front panel, keypad, paddle latches, paddles, cables, strain reliefs, and battery seating.
Expected outcome: Unit is clean, intact, undamaged, and free of excessive wear. Cables should have no cracks, cuts, exposed wires, or damaged strain reliefs.
Remove from service if: Cracked housing, damaged connectors, exposed wires, failed paddle latches, damaged paddles, battery seating issue, or any condition that could affect safe clinical use.
3. Front Panel Button and Recorder Check
Action: Power the unit on in Monitor / ON mode. Verify startup tones, lead selection, ECG size changes, alarm suspend function, recorder operation, check-recorder message with door open, AC power LED, battery LED, energy selection, charge function, and shock button operation.
Expected outcome: Controls respond correctly, recorder prints correctly, messages clear properly, and energy selection/charge/discharge functions operate as expected.
4. ECG Lead Test
Action: Using the defibrillator analyzer and applicable ECG cables, verify 3-lead and/or 5-lead ECG operation. Disconnect and reconnect each lead to verify lead-off detection.
Expected outcome: ECG lead-off messages display and clear appropriately within expected behavior.
5. Optional Power Supply / Charger Test
Action: Perform only if included in facility PM workflow. Follow the manual setup using DC power supply, resistors, DMM, and battery/charger fixture.
Expected outcomes listed in the manual include:
- Unit should not power on at 7 V.
- Unit should power on at 10.8 V with no LOW BATTERY message.
- LOW BATTERY should display at 10.5 V within 30 seconds.
- REPLACE BATTERY should display at 10.2 V within 30 seconds.
- System current should be within manual-listed limits.
- Charger voltage should meet manual-listed ranges.
- 200 J charge time should be 3–10 seconds during charger test.
6. Leakage Current Test
Action: Perform leakage current testing using the electrical safety analyzer / leakage tester per the tester manufacturer’s instructions. Repeat with applicable accessories: OneStep cable, external paddles, internal paddles, and anterior/posterior paddles when applicable.
Manufacturer-listed maximum leakage acceptance limits:
- Lead ECG: 10 µA normal condition, 50 µA single fault condition
- Hands-free electrodes / paddles / MFE: 10 µA normal condition, 100 µA single fault condition
- Earth: 500 µA normal condition, 1000 µA single fault condition
The manual notes the single fault condition as AC mains on applied part.
7. Paddles Test
Action: If paddles are used, verify paddle energy select buttons, recorder button, charge button, individual shock-button behavior, and dual-button discharge.
Expected outcome: Energy selection changes correctly, recorder starts/stops, unit charges to selected energy, individual paddle shock buttons do not discharge alone, and both shock buttons together discharge properly.
8. Heart Rate Display Test
Action: Connect ECG cable to analyzer. Set simulated rhythm to 120 BPM.
Expected outcome: Heart rate displays 120 ± 2 BPM.
9. Calibration Pulse / Strip Chart Test
Action: Activate the calibration signal and print strip chart.
Expected outcome: Strip chart displays a 300 ppm signal with amplitude 10 mm ± 1 mm.
10. Notch Filter Test
Action: Use the analyzer to input 50 Hz or 60 Hz sine wave depending on unit configuration.
Expected outcome: Waveform amplitude on strip chart is less than 1.5 mm.
11. Heart Rate Alarm Test
Action: Verify alarm menu navigation, enable/disable function, lead-off alarm, high heart rate alarm, alarm suspend, and alarm disable behavior.
Expected outcome: Alarm states, bell symbol behavior, highlighted heart rate values, and alarm tones respond as expected.
12. Defibrillator Self-Test
Action: Perform the 30 J self-test using the OneStep cable and test port as described by the manual.
Expected outcome: Unit displays 30J TEST OK. Printed strip should indicate 30 JOULES TEST OK, with test current and impedance within the manual-listed expected range.
Remove from service if: Self-test fails, unit does not charge/discharge correctly, or unexpected defib fault appears.
13. Synchronized Cardioversion Test
Action: Connect the OneStep cable to the defibrillator analyzer using the adapter. Input 1 mV ECG at 60 BPM. Enable sync and deliver synchronized shock.
Expected outcome: Sync markers appear over R-waves. Sync delay is less than 60 ms.
14. Remote Sync Cardioversion Test — If Applicable
Action: If the unit is configured for remote sync, test the complete system with the remote monitor connected to the Sync In/Marker Out connector.
Expected outcome: Remote sync displays properly and sync delay is less than 60 ms.
15. Shock Output Test
Action: Using a defibrillator analyzer and fully charged battery, test output at manufacturer-listed energy settings.
Expected outcomes:
- 5 J selected: delivers 3–7 J
- 50 J selected: delivers 46–62 J
- 100 J selected: delivers 93–125 J
- 200 J selected: delivers 196–264 J
- 200 J charge time: 1.0–7.0 seconds
- Patient current: 23.9–25.9 A
- Defib impedance: 46–54 Ω
- Unit holds energy for 60 seconds on manual units, 15 seconds on AED units, then internally discharges without fault messages.
16. Summary Report Test
Action: Erase prior report data if appropriate, deliver a shock into analyzer, add a code marker, power cycle, and print the summary report.
Expected outcome: Summary report prints with correct date, time, shock delivered, and code marker event.
17. Advisory Message Test — Manual / Advisory Units
Action: Using analyzer rhythm simulations, verify advisory behavior with VF and NSR.
Expected outcome: VF results in shock-advised sequence. NSR results in no-shock-advised behavior.
18. Pacer Test — If Installed
Action: Using external pacer analyzer or equivalent, verify pad contact messages, output accuracy, rate accuracy, pulse width, demand behavior, asynchronous pacing, recorder markers, and 4:1 behavior.
Expected outcomes include:
- 40 mA output: 40 mA ± 5 mA
- 80 mA output: 80 mA ± 5 mA
- 120 mA output: 120 mA ± 6 mA
- 140 mA output: 140 mA ± 7 mA
- Pulse width: 40 ms ± 2 ms
- 180 ppm setting: 177–183 ppm
- 30 ppm setting: 29–31 ppm
19. SpO₂ Monitor Test — If Installed
Action: Use compatible SpO₂ simulator, patient cable, and sensor. Verify SpO₂ value, pulse rate, pleth display, recorder output, and cable disconnect response.
Expected outcomes include:
- 98% simulator setting: R Series displays 98 ± 1%
- 230 BPM setting: R Series displays 226–234 BPM
- 50 BPM setting: R Series displays 46–54 BPM
- 72% setting: R Series displays 69–74%
20. EtCO₂ Monitor Test — If Installed
Action: Connect CAPNOSTAT 5 cable and airway adapter. Allow warm-up, zero the sensor, display waveform, and verify capnogram response.
Expected outcome: Warm-up completes, ZERO DONE displays, flat baseline appears, and capnogram waveform appears when breathing through adapter.
21. Barometric Pressure Calibration Check — If EtCO₂ Installed
Action: Enter EtCO₂ calibration screen and compare/set barometric pressure using local barometric pressure converted to mmHg if needed.
Expected outcome: Barometric pressure value is stored and calibration screen returns normally.
22. CO₂ Accuracy Check — If EtCO₂ Installed
Action: Use calibration gas and regulator. Set gas temperature and gas balance as required.
Expected outcome: With 5% CO₂ calibration gas, displayed value is 5% ± 0.2%.
23. NIBP Volume Leak Test — If Installed
Action: Use NIBP simulator. Enter NIBP service mode, close valves, apply 200 mmHg, and start volume leak test within 30 seconds.
Expected outcome:
- Volume leak reading <55 = pass.
- Volume leak reading >56 = fail.
- No volume leak reading but stable pressure at or above 200 mmHg = pass.
- If using Fluke Biomedical CuffLink, manual notes <10 as pass and >10 as fail.
24. NIBP Transducer Calibration Test — If Installed
Action: Perform two-point calibration at 0 mmHg and 250 mmHg, then verify a different pressure value.
Expected outcome:
- 0 mmHg calibration field displays PASS.
- 250 mmHg calibration field displays PASS.
- Verification pressure reads within ±3 mmHg of simulator pressure.
The manual warns that NIBP transducer calibration can affect clinical NIBP readings and should be followed by the NIBP Monitor Test.
25. NIBP Monitor Test — If Installed
Action: Use NIBP simulator with manual-listed values and facility-established simulator offset when applicable.
Expected outcome using BP Pump 2 setup:
- Systolic: 126 ± 5 mmHg
- Diastolic: 84 ± 5 mmHg
- Mean: 98 ± 5 mmHg
- Pulse rate: 77–85 BPM
Electrical Safety Testing
Perform leakage current testing using the facility’s electrical safety analyzer and the manufacturer-listed leakage acceptance limits.
Ground resistance measured in ohms, expected ≤ 0.5 Ω or facility/NFPA 99 criteria if different.
- Lead ECG leakage: 10 µA normal condition, 50 µA single fault condition
- Hands-free electrodes / paddles / MFE leakage: 10 µA normal condition, 100 µA single fault condition
- Earth leakage: 500 µA normal condition, 1000 µA single fault condition
Touch/chassis leakage measured in microamps:
Facility limit used =
Result =
Pass / Fail =
Pass / Fail Criteria
PM Pass Criteria
Pass the PM only if:
- Physical inspection passes.
- Unit powers on and controls function correctly.
- Recorder operates correctly.
- ECG lead detection and heart rate display pass.
- Leakage current results are within manufacturer-listed and facility limits.
- Defibrillator self-test passes.
- Shock output values are within manufacturer-listed ranges.
- Sync delay is less than 60 ms.
- Alarms function correctly.
- Pacing passes if installed.
- SpO₂ passes if installed.
- EtCO₂ passes if installed.
- NIBP leak, calibration, and monitor tests pass if installed.
- No unresolved fault messages remain.
Remove From Service Criteria
Remove from service if:
- Unit fails defibrillator output, charge, self-test, or sync testing.
- Leakage current exceeds limits.
- Pacer output/rate/pulse width fails on units with pacing.
- NIBP calibration or monitor test fails and cannot be corrected.
- EtCO₂ or SpO₂ option fails where clinically used.
- Housing, connectors, cables, paddles, or battery well are damaged.
- Unit displays unresolved fault messages.
- Any condition could affect safe defibrillation, pacing, monitoring, or transport use.
Copy/Paste CMMS Work Order Procedure
Use the manufacturer service manual, facility policy, and approved AEM program to determine the final PM interval and return-to-service requirements.