On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide assists Clinical Engineering in troubleshooting a Skytron 6500 Series surgical table when the table powers on but movement functions (raising, lowering, tilting, or lateral adjustments) do not operate. Common symptoms include:
- Table powers on with indicator lights functioning
- No response from hand or foot controls
- Audible motor or hydraulic sounds may be absent or abnormal
This issue is commonly noticed during room setup, patient positioning, or after the table has been moved for cleaning or turnover. If the table cannot raise, lower, tilt, or articulate reliably, it should not be used for a procedure until the movement issue is verified and corrected.
Focus is on external, verifiable checks before assuming internal mechanical or electrical failure.
Step-by-Step Troubleshooting
Verify Power Source
- Confirm the table is plugged into a live, dedicated outlet.
- Check for tripped breakers or blown fuses.
Why: Even if the table shows partial power (lights), movement motors require full voltage.
Check Main Power Switch
- Ensure the main power switch is in the “ON” position.
- Some Skytron models have separate switches for power and motion enable.
Why: Movement functions may be disabled if the motion enable is off.
Inspect Hand and Foot Controls
- Ensure all buttons are functioning and not stuck.
- Verify connection cables are fully seated.
- Swap with a known good pendant if available.
If controls appear unresponsive, see Skytron 6500 Series - Hand Control or Pendant Failure for focused troubleshooting of pendant-related issues.
On surgical tables, a damaged pendant cable, loose connector, or stuck button can make the entire table appear locked out even when the table itself has power.
Why: Faulty or disconnected control units often mimic motor failure.
Observe Indicator Lights or Error Codes
- Check display panel for error codes or flashing indicators.
- Document any patterns for repair escalation.
Why: Error codes can indicate motor drive or safety interlock issues.
Check Safety Interlocks
- Verify table brakes are disengaged and base locks are not activated.
- Ensure any side rails or attachments are correctly positioned.
Also confirm the table is not loaded with accessories, attachments, or positioning equipment that could be interfering with normal motion or triggering a safety inhibit condition.
Why: Many Skytron tables will inhibit movement if safety conditions are not met.
Inspect Power Cables and Connectors
- Look for loose or damaged cables, including those connecting the power module to motors.
Why: Intermittent or damaged wiring can prevent movement.
Test Motion Functions Individually
- Attempt raising/lowering, tilt, and lateral movements one at a time.
- Note any movement or sounds.
If the table moves inconsistently or does not hold position, refer to Skytron 6500 Series - Table Drift or Position Instability for additional guidance.
Why: Isolates which circuits or motors are affected.
If the Problem Persists
All external and easily verifiable causes have been ruled out. The issue is likely internal (motor controller, drive circuit, hydraulic actuator).
Action:
- Remove the table from service
- Label “Out of Service”
- Send to manufacturer repair or qualified bench evaluation
Knowing when to escalate prevents unsafe troubleshooting and protects the table from further damage.
Clinical Use Tip
- Never attempt troubleshooting while a patient is on the table.
- Move the patient to an alternative table first.
- Only attempt troubleshooting when the table is unoccupied and safely powered down.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“Skytron 6500 Series surgical table powers on but will not raise, lower, tilt, or respond to movement controls during room setup.”
Cause
What was observed during troubleshooting.
Example:
“Verified outlet power, main power switch, pendant connection, foot control connection, and safety interlocks. Table remained inoperative with no movement response from available controls.”
Resolution
What action was taken.
Example:
“Table removed from service and labeled Out of Service. Unit referred for qualified service evaluation due to suspected internal motion control, motor drive, or hydraulic actuator fault.”
Helpful Details to Include
- Outlet verified live
- Control cables checked or swapped
- Brakes and interlocks verified disengaged
- Indicator lights normal or error codes documented
- No unusual noises, heat, or odor observed
- Table removed from service for repair
Final Thought
Patient safety is the priority. Logical, stepwise troubleshooting helps identify external causes efficiently. Escalation to repair is appropriate when external checks fail. Document all observations clearly for continuity of care and future reference.
That is successful troubleshooting.